Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Automation has dramatically changed both the analytical and nonanalytical aspects of clinical laboratory operations. Automation of laboratory test procedures began more than 50 years ago, but nonanalytical automation—including conveyor systems, interfaced analyzers, and automated specimen processing and storage—began in earnest in the 1990s. Today there is a wide selection of automation options designed to improve the quality, throughput, and efficiency of laboratory testing.
This chapter covers automation from both nonanalytical and analytical perspectives. Historical contexts are provided. The discussion of preanalytical automation includes a review of labeling, barcoding, and portable wireless labeling systems along with the use of pneumatic tube systems and mobile robots for transport of specimens. Single-function robotic systems and multifunction systems for specimen processing are discussed. Some of these systems have pre- and postanalytical capabilities. Total laboratory automation (TLA) systems are discussed extensively. Several TLA systems include postanalytical functions; thus the chapter also addresses storage and retrieval systems. The second half of the chapter discusses automation from the analytical perspective, including specimen and reagent handling on analytical instrumentation, as well as common measurement approaches used by automated analyzers. The chapter concludes with area-specific considerations of how analytical automation has impacted all subdisciplines of laboratory medicine.
The term automation has been applied in clinical chemistry to describe the process whereby an analytical instrument performs many tests with only minimal involvement of an analyst. However, the term also applies to the automation of nonanalytical processes (preanalytical and postanalytical) that have become very important to the performance of laboratories over the past two decades. This chapter covers both nonanalytical and analytical automation in clinical laboratories.
The availability of automation systems and automated instruments enables laboratories to process much larger workloads without comparable increases in staff. The evolution of automation in clinical laboratories has paralleled that in the manufacturing industry, progressing from fixed automation, whereby an instrument performs a repetitive task by itself, to programmable automation, which allows the instrument to perform a variety of different tasks. Intelligent automation also has been introduced into some individual instruments or systems to allow them to self-monitor and respond appropriately to changing conditions.
One benefit of automation is a reduction in the variability of results and errors of analysis through the elimination of tasks that are repetitive and monotonous for most individuals. The improved reproducibility gained by automation has led to a significant improvement in the quality of laboratory tests.
Many small laboratories now have consolidated into larger, more efficient entities in response to market trends involving cost reduction. The drive to automate these mega-laboratories has led to new avenues in laboratory automation. No longer is automation simply being used to assist the laboratory technologist in test performance; it now includes (1) processing and transport of specimens, (2) loading of specimens into automated analyzers, (3) assessment of the results of the tests performed, and (4) storage of specimens. We believe that automating these additional functions is crucial to the future prosperity of clinical laboratories. ,
Clinical laboratory automation as we know it today owes much to the pioneering efforts of Masahide Sasaki, MD, PhD, professor and director of the clinical laboratory at the Kochi Medical School in Kochi, Japan. The authors are indebted to Dr. Robin Felder for the details of Dr. Sasaki’s work described in this paragraph. Beginning in 1981, Sasaki trained his medical technologists to assemble conveyor belts, construct electronic boards, and program robots, thus leading to the creation of the world’s first automated laboratory. This amazing development was first revealed to Western audiences at the annual meeting of the American Association for Clinical Chemistry in Atlanta, Georgia, in July 1989. In front of a spellbound standing-room-only audience of 900 attendees, Sasaki’s 45-minute video demonstrated new possibilities for error reduction, cost containment, and unprecedented turn-around times. This led to a steady stream of visitors to his laboratory and collaborations with major industrial partners that brought commercial automation products to clinical laboratories throughout Japan.
Much of the automation we see in our laboratories today traces its lineage to these early automation systems in Japan. Thousands of laboratories worldwide now have either total laboratory automation (TLA) systems or automated systems that perform specific functions. Automation continues to grow as it is widely recognized as a principal means of eliminating errors; improving quality; and reducing labor, costs, and turn-around time.
Significant progress has been made in automating pre- and postanalytical activities and integrating these operations with analytical systems. Preanalytical automated systems range from robotics that perform a single function such as aliquoting or sorting (also called task-targeted automation systems ) to more complex workstations that perform several preanalytical functions. Several options are also available as modules for TLA systems, which are discussed later in this section on nonanalytical automation.
Preanalytical automation begins with automation of operations in the specimen processing area where specimens are (1) identified, (2) labeled, (3) scheduled for analysis, (4) centrifuged, and (5) sorted. After specimens are processed, they are transported to appropriate workstations in the laboratory, either manually or using conveyors, where they can then be analyzed with minimal or no human intervention. Rules-based expert system software (1) assists with the review of laboratory results by automatically releasing results that have no associated problems and (2) identifies any problematic results to bring to the attention of trained medical technologists. All specimens are catalogued after analysis and stored in a central storage facility, which may include automated storage and retrieval functions. The following sections cover various aspects of nonanalytical processes with descriptions of automation for these activities.
Typically, the identifying link (identifier) between the patient and specimen is noted at the patient’s bedside. Maintenance of this connection throughout (1) transport of the specimen to the laboratory, (2) subsequent specimen analysis, and (3) preparation of a report is essential. Several technologies are available for automatic identification and data collection purposes ( Box 29.1 ). In practice, automatic identification includes only technologies that electronically detect a unique characteristic or unique data string associated with a physical object. For example, identifiers such as (1) serial number, (2) part number, (3) color, (4) manufacturer, (5) patient name, (6) medical record number, and (7) accession number have been used to identify an object or patient through the use of electronic data processing. In a clinical laboratory, labeling with a bar code has become the technology of choice for purposes of automatic identification.
Bar coding
Optical character recognition
Magnetic stripe and magnetic ink character recognition
Voice identification
Radiofrequency identification (RFID)
Touch screens
Light pens
Hand print tablets
Optical mark readers
In many laboratory information systems (LIS), electronic entry of a test order in the laboratory or at a nursing station for a uniquely identified patient generates a specimen label bearing a unique laboratory accession number. A record is established that remains incomplete until a result (or a set of results) is entered into the computer against the accession number. The unique label is affixed to the specimen collection container when the specimen is obtained and the patient is properly identified. This may be at the bedside using preprinted labels carried from the laboratory or using a portable label printer connected wirelessly through the network to the laboratory computer. Many companies now provide these portable bar code labeling systems, and many hospitals are now using them. Proper alignment of the label on a specimen tube can be critical for correct processing of bar-coded labels, although newer imaging systems can be more tolerant. The Clinical and Laboratory Standards Institute (CLSI) published a standard in 2011 ( Specimen Labels: Content and Location, Fonts, and Label Orientation ) that was intended to standardize specimen labels in clinical laboratories to reduce errors such as patient misidentifications. The focus of this standard was not the bar code but rather the human-readable content of the label, particularly the location and appearance of the patient name.
An electronic or paper requisition form typically initiates specimen acquisition. Labels to identify the patient and the date and time of collection are typically applied to specimen containers and tubes at the time of collection. Many systems also assign a laboratory accession number at this time. Alternatively, accession numbers may be assigned upon arrival of the sample in the laboratory during a log-in procedure that includes date, time, and condition of the specimen.
After receipt by the laboratory, specimens undergo various technical handling processes. Some automated analyzers sample directly from the original collection tube, simultaneously reading the accession number from the bar code label on the tube; other analyzers require an aliquot of the serum or plasma in the original tube. For processes requiring physical removal of serum or plasma from the original tube, secondary labels bearing essential information from the original label must be affixed to any secondary tubes created. Secondary bar code labels, if necessary, may be generated at the time of accessioning or at the time the aliquots are prepared. This aliquoting step has been automated with different systems.
A major advance in the automation of specimen identification in clinical laboratories is the incorporation of bar-coding technology into analytical systems. A bar coding system consists of a bar code printer and a bar code reader, or scanner. In clinical laboratories, one-dimensional or linear bar code systems have historically been used. CLSI published a standard (AUTO02-A) in 2000 and updated it (AUTO02-A2) in 2003, specifying that clinical laboratories should only use the Code 128B symbology on specimen labels and phase out less reliable symbologies such as Code 39, Codabar, and Code I2of5. These latter symbologies could still be used by laboratories for other purposes such as reagents, inventory items, documents, and so on. However, as this chapter was in preparation, CLSI was developing a new standard (AUTO14), expected to be published in 2021, for the use of two-dimensional bar codes in clinical and anatomic pathology laboratories. Two-dimensional bar codes offer the prospect of encoding more data in smaller formats and having greater reliability than one-dimensional bar codes.
A one-dimensional bar code is an array of rectangular bars and spaces arranged in a predetermined pattern according to unambiguous rules to represent elements of data referred to as characters . A bar code is transferred and affixed to an object by a bar code label that carries the bar code and, optionally, other noncoded human readable information. Symbology is the term used to describe the rules specifying the way data are encoded into the bars and spaces. The width of the bars and spaces, as well as the number of each, is determined by a specification for that symbology. Different combinations of bars and spaces represent different characters. When a bar code scanner is passed over the bar code, the light beam from the scanner is absorbed by the dark bars and is not reflected; the beam is reflected by the light spaces. A photocell detector in the scanner receives the reflected light and converts that light into an electrical signal that is then digitized. Alternatively, for those systems that acquire images, digital data is extracted directly from the width of the dark and light bars in the image. A one-dimensional bar code is “vertically redundant” in that the same information is repeated vertically—the heights of the bars can be truncated without any loss of information. Vertical redundancy typically allows a symbol with printing defects, such as spots or voids, to still be read.
In practice, a bar code label (often generated by the LIS and bearing the sample accession number) is placed onto the specimen container and is subsequently “read” by one or more bar code readers placed at key positions in the analytical sequence. Examples of these key positions are a handheld bar code reader in a laboratory section, a task-targeted automation system, the loading unit for a TLA system, and an automated analyzer. The resultant identifying and ancillary information is then transferred to and processed by the system software. Generating bar codes prior to specimen acquisition with identification at a patient’s bedside ensures greater integrity of the specimen’s identity throughout these processes. Chapter 4 provides greater detail on specimen collection and handling.
Unequivocal positive identification of each specimen is achieved in analyzers with bar code readers. Advantages of the use of bar code labels include the following:
Elimination of work lists for the system
Avoidance of mistakes made in the placement of tubes in the analyzer or during sampling
Avoidance of the need for analysis of specimens in a defined sequence
Decrease in identification errors.
A best practice to reduce errors caused by bar code misreads is to set the bar code readers on analyzers and automation systems to only read the symbology expected on the specimen labels (e.g., Code 128B). This prevents reading errors in the event that a second bar code of a different symbology might be visible on an underlying label (e.g., a tube vendor’s bar code that indicates the tube type and its additive).
Many opportunities arise for the mismatch of specimens and results. The risks begin at the bedside and are compounded with each processing step a specimen undergoes between collection from the patient and analysis by the instrument. The risks are particularly great when hand transcription is invoked for accessioning, labeling and relabeling, and creation of lists of samples for analysis in one batch or instrument, also referred to as “load lists.” An incorrect accession number, one in which the digits are transposed, or a load list with transposed accession numbers may cause test results to be attributed to the wrong patient. An additional hazard exists when specimens must be inserted into certain positions in the loading zone defined by a load list. Human misreading of the specimen label or the loading list may cause misplacement of specimens, calibrators, or controls. Automatic reading of bar code labels reduces the error rate from 1 in 300 characters (for human entry) to about 1 in 1 million characters barring (1) imperfections in printed bar codes, (2) improper bar code scanner resolution, or (3) skewed orientation of bar code labels on containers, all of which can result in read errors. Reading errors include the inability to read bar codes requiring manual intervention, or more critically, incorrect reading that incorporates misinformation without warning. For example, Snyder et al. reported patient identification errors caused by misreading of one-dimensional bar codes in point-of-care (POC) testing using patient wristbands. An editorial accompanying that report further explained the problems that may exist in the use of linear bar codes. This concern about the overall reliability of linear bar codes has provided the impetus for CLSI to develop a standard for the use of two-dimensional bar codes.
The clotting of blood in specimen collection tubes, their subsequent centrifugation, and the transfer of serum to secondary tubes require a finite time to complete. If performed manually, the process may result in a delay in the preparation of specimens for analysis. To eliminate the problems associated with specimen preparation, systems are being developed to automate this process.
When whole blood is used in an assay system, specimen preparation time is essentially eliminated. Automated or semiautomated ion-selective electrodes, which measure ion activity in whole blood rather than ion concentration, have been incorporated into automated systems to provide certain test results within minutes of collecting a specimen. Comparable electrode-based methods can be used for glucose, creatinine, and lactate. Another simple approach involves manual or automated application of whole blood to dry reagent films and visual or instrumental observation of a quantitative change (see Chapter 30 ).
Automated methods are often used to deliver specimens to the laboratory instead of more manual methods (e.g., phlebotomist transport, courier service, nurse delivery). Pneumatic tube systems have been used in many hospitals, but mobile robots can also transport specimens and other laboratory supplies within a laboratory or to the laboratory from other locations in the facility.
Pneumatic tube systems provide rapid specimen transport and are reliable when installed as point-to-point services. However, when switching mechanisms are introduced to allow carriers (the bullet-shaped containers used to hold specimens) to be sent to various locations, mechanical problems may occur and may cause misrouting. In addition, close attention to the design of the pneumatic tube system is necessary to prevent hemolysis of the specimen. Avoidance of sudden accelerations and decelerations and the use of proper packing material inside the carriers can help to minimize hemolysis. Smart phones have also been used to monitor acceleration forces and turbulence experienced by specimens. Pneumatic tube systems should be validated prior to use in order to ensure that analytical results are not impacted by hemolysis.
Automated guided vehicles (AGVs), also called mobile robots, have been used successfully to transport laboratory specimens both within a laboratory and outside a central laboratory. They are easily adapted to carry various sizes and shapes of specimen containers and are reprogrammable with changes in laboratory geometry. In addition, in a busy laboratory setting, delivery of specimens to laboratory benches by a mobile robot can be more frequent than human pickup and is cost-effective. Inexpensive models follow a line on the floor, but others have more sophisticated guidance systems. Their limitations include the need to batch specimens for greater efficiency and, in most cases, the requirement for laboratory personnel to place specimens onto or remove specimens from the mobile robot at each stopping place. Some mobile robots have been integrated with robotic systems that automate loading and unloading of specimens; others initiate an audible or visual signal of their arrival at a specified station so that employees are able to load or unload the specimens being transported. Box 29.2 lists several vendors that provide mobile robot systems for clinical laboratories.
Aethon: https://aethon.com/
AGVE : https://www.agvegroup.com/
Oceaneering: https://www.oceaneering.com/automated-guided-vehicles/
Swisslog: https://www.swisslog.com/
Some laboratories have gained efficiency, improved quality, and lowered turn-around time by implementing an automated system that performs a single task. This automation is also referred to as task-targeted automation . Examples of such automation include automated centrifuges, decappers, recappers, aliquotters, and sorters. Such systems began to be developed in the 1990s and early 2000s, when TLA systems were relatively expensive and thought to be only affordable to reference laboratories and the largest hospital laboratories. Today these systems continue to be developed and implemented because they offer affordable solutions for specific customer needs. Box 29.3 lists a number of vendors offering one or more single-function workstations. Fig. 29.1 is an example of a sorter, one of the different types of single-function workstations that are available. However, a laboratory could use a sorter for both preanalytical sorting for different testing areas and postanalytical sorting for archival storage.
Hettich: https://www.hettweb.com
IDS Co., Ltd.: http://www.idsma.com/en/
m-u-t automation: https://www.mut-group.com/en/laboratory-automation
Roche Diagnostics: https://diagnostics.roche.com
Sarstedt: https://www.sarstedt.com
Yaskawa America, Inc. Motoman Robotics Division: https://www.motoman.com/applications/clinical-lab
Although the manual operations carried out in a specimen processing area may look simple, considerable complexity underlies them. Consequently, specimen processing has been one of the most difficult areas of the clinical laboratory to automate. Each specimen passing through a specimen processing area has to undergo a series of operations, including (1) receiving the specimen; (2) inspecting it for appropriateness (labeling, container type, temperature, and quantity of specimen); (3) logging the specimens into the LIS; (4) recording the date and time of collection and phlebotomist ID, if not already entered; (5) labeling the specimen with an accession number if not already there; and (6) separating urgent and stat specimens from routine specimens. Also, specimens may have to be sorted for centrifugation and aliquoting and may have to be sorted or otherwise prepared for the appropriate laboratory station.
An example of a stand-alone specimen processing system is shown in Fig. 29.2 . Such systems place processed specimens into racks that must be transported manually to the testing areas, with some exceptions. Some of these systems are approximately the size of a large, automated analyzer; others may be slightly larger. They may be a good choice for laboratories (1) with daily workloads of 500 to 1500 specimens, (2) with space limitations, or (3) that desire ease of use with different analyzers from different vendors. Some laboratories may choose to use multiples of a stand-alone specimen processing system based on higher daily volumes or to automate postanalytical archiving in addition to preanalytical specimen processing.
Box 29.4 lists vendors offering multifunction or specimen processing workstations. Although there may be some variation in the functions that are included, these systems will typically (1) receive incoming specimens, (2) sort, (3) decap, (4) aliquot, and (5) label aliquot specimen containers with bar codes. All may be interfaced to the laboratory’s LIS. Some systems may also include automated centrifugation. Some of the systems sort into instrument-specific racks for analyzers from a number of different vendors. In addition, some users apply these systems to aliquot and sort reference or “send-out” testing, saving considerable time in locating the original specimens after testing in their own laboratory.
Abbott Diagnostics: https://www.corelaboratory.abbott/int/en/home
Aim Lab Automation Technologies: https://www.aimlab.com
Beckman Coulter: https://www.beckmancoulter.com
Becton Dickenson : https://www.bd.com/en-us/offerings/capabilities/lab-automation
COPAN Diagnostics: https://www.copanusa.com/
IDS Co., Ltd.: http://www.idsma.com/en/
Roche Diagnostics: https://diagnostics.roche.com
Sarstedt: https://www.sarstedt.com
Yaskawa America, Inc. Motoman Robotics Division: https://www.motoman.com/applications/clinical-lab
Several manufacturers offer integrated or modular automation systems that use conveyor belts to connect preanalytical specimen processing and other functions directly to analyzers. Such systems may also include functions such as postanalytical storage or sorting of specimens or aliquots to be transported to low-volume testing areas. A list of automation vendors offering TLA systems can be found in Box 29.5 ; Fig. 29.3 illustrates a large TLA system with a variety of connected nonanalytical and analytical modules.
Abbott Diagnostics: https://www.corelaboratory.abbott/int/en/home
Beckman Coulter: https://www.beckmancoulter.com
IDS Co., Ltd.: http://www.idsma.com/en/
Inpeco: https://www.inpeco.com
Ortho-Clinical Diagnostics: https://www.orthoclinicaldiagnostics.com
Roche Diagnostics: https://diagnostics.roche.com
Siemens Healthcare Diagnostics: https://www.siemens-healthineers.com/laboratory-automation
Sysmex America: https://www.sysmex.com/us
Yaskawa America, Inc. Motoman Robotics Division: https://www.motoman.com/applications/clinical-lab
In addition to the nonanalytical functions listed in the preceding section, TLA systems typically add (1) conveyor transport; (2) interfacing to automated analyzers; (3) more sophisticated process control; and, in some cases, (4) a specimen storage and retrieval system. All systems are of modular design, allowing the customer to choose the modules or features that should be included. Some systems use an open design, which permits interfaces to analyzers from a variety of vendors, but other systems are of a closed design and are interfaced only to the vendor’s own (or a limited number of other) analyzers. Closed systems typically do not have process control software that is independent of the instruments or system; rather, the automation process control is integrated to work with the vendor’s analyzers.
To achieve maximum effectiveness of an automation system, process control software should be able to read the specimen’s ID bar code and obtain information from the LIS about specimen type and ordered tests. This process control software is usually referred to as the Laboratory Automation System (LAS), not to be confused with the actual hardware of the laboratory automation. Readers are referred to the CLSI standard on automation communications (AUTO03-A2) for further information on the LAS. The software should then determine the processes the specimen requires and the exact route or course of action for each specimen. It should be able to (1) calculate the number of aliquots and the proper volume for each specimen depending on the tests requested, (2) route the specimens to preanalytical processing equipment and then to analyzers, (3) recap the specimens after testing, and (4) retain the specimens for automatic recall. The software should be able to monitor analyzers for in-control production status and automatically make decisions if a test is not available. Specimen integrity for hemolysis, icterus, and lipemia should be performed. Having the automation system do specimen integrity checking, if possible, may be more efficient than having it conducted farther downstream in analyzers. However, automatic spectrophotometric methods are available on many modern analyzers. Rules-based decisions from analysis at multiple wavelengths may correct the effects of interferents for some assays and alert the laboratory on others. Finally, most process control software should include (1) “autoverification,” which is validation of analyzer results by making rules-based decisions that flag exceptions for technologist review, and (2) “autoretrieval” of specimens for repeat, reflex, and dilution testing. Assistance with designing an autoverification system can be found in the CLSI Guideline AUTO10-A.
Although most of these systems are restricted to handling specific types of specimen containers, they are capable of processing much of the daily workload of a large clinical laboratory. A few laboratories with daily workloads as low as 600 to 800 specimen tubes have justified these systems because of a shortage of technical help, but typically these systems are designed for laboratories with workloads of 1000 to 25,000 specimens per day. In addition to process control software and an LIS interface, each of these systems incorporates some or all of the following components:
Specimen input area: This is a holding area where bar code labeled specimens are introduced into the system.
Bar code reading stations: Multiple bar code readers are placed at critical locations in the processing system to track specimens and provide information for their proper routing to various stations. Alternately, bar code readers may only be required at limited locations in the system if each transport carrier (puck) has a unique ID such as a radiofrequency identification (RFID) chip, and the specimen’s bar code ID is “married” to the puck’s ID when first placed in the puck.
Transport system: Segments of a conveyor belt line that move specimens to the appropriate location.
A high-level device to sort or route specimens: A device that separates specimens based on processing requirements or order code and passes them to the transport system or to a system using racks. A high-level sorter is often used to separate specimens that require centrifugation or other processing steps from specimens that do not or to route specimens into completely different pathways within the total automation system.
Automated centrifuge: An area of the specimen processor in which specimens requiring centrifugation are removed from the conveyor belt, introduced into a centrifuge that is automatically balanced, centrifuged (refrigerated or at room temperature), and then removed from the centrifuge and placed back on the transport system.
Level detection and evaluation of specimen adequacy (specimen integrity): An area where sensors are used to evaluate the volume of specimen in each specimen container and to look for the presence of hemolysis, lipemia, or icterus. These sensors can be located on one of the chemistry analyzers on the system and data shared across the laboratory or can be located in one dedicated area.
Decapper station: An area or device in the automated system where specimen caps or stoppers are automatically removed and discarded into a waste container.
Recapper station: An area or device in the automated system where specimen tubes are automatically recapped with new screw caps, stoppers, or an alternate air-tight closure.
Aliquotter: A machine that aspirates appropriately sized aliquots from each original specimen container and places them into bar-coded secondary specimen containers for sorting and transport to multiple analytical workstations. The LAS generally instructs the aliquotter as to how many aliquots of what volumes are required based on the bar code ID of the specimen.
Physical interfaces to automated analyzers: A direct physical connection to an automated analyzer that permits the analyzer’s sampling probe to aspirate directly from an open specimen container while the container is still on the conveyor or that may robotically lift the container from the conveyor and place it onto the analyzer. Whereas some automation systems interface only to their own brand of analyzers or to a limited number of systems, other automation systems use a so-called open design that complies with CLSI standards and permits interfaces to a variety of automated analyzers.
Sorter: An automated sorter to sort specimens not going to a conveyor-interfaced analyzer or workstation. Such a sorter typically sorts into 30 to 100 different sort groups in racks or carriers. In some systems, the racks may be specific to certain analyzers for convenience. Some automation systems may use a sorter as a temporary holding area for specimens that may require a repeat test (see take-out stations).
Take-out stations: Temporary storage areas for specimens before or after analysis. The take-out station may be the same as the sorter described earlier, where specimens are sorted for manual delivery. However, it may also serve as a holding area (stockyard) for specimens awaiting autoverification of results in case a repeat test is required.
Storage and retrieval system: This unit may serve the same function as the take-out station or stockyard—that of holding specimens after analysis in case a specimen is necessary for a repeat test—but it has one major difference. These units are typically refrigerated and hold many more specimens (3000 to 30,000) than the typical take-out station or stockyard. Depending on daily workloads, the laboratory may be able to retain up to 1 week’s worth of specimens for possible repeat or additional tests. Specimen containers are loaded and retrieved with a robot. Some storage systems include automated specimen discard directly into biohazard waste bins after a predetermined storage duration has been reached.
Analytical automation has impacted all areas of laboratory medicine.
Along with automating the testing phase, analyzers may incorporate specimen handling and reagent management, as well as quality control and result review.
Automated processes, including liquid-level sensing, clot detection, closed-container sampling, and instrument autodilution, help to improve the efficiency of laboratory testing.
Analytical systems can be organized into clusters or total laboratory automation solutions, allowing operators to control multiple instruments.
One main feature of integrated TLA systems is the use of conveyor belts. Ordinary industrial conveyor belts have been used successfully when only transportation is required. However, in the past, when conveyors were integrated with other robotic systems to automate pre- or postanalytical functions, this technology had difficulty handling the large variety of specimen containers found in the clinical laboratory. To increase the variety of types of specimen containers that are carried on a conveyor belt system, specimens are placed into specially designed carriers that fit on the conveyor belt line. Sometimes known as “pucks” or “racks” (depending on whether they carry individual specimens or groups of specimens), the carriers have receptacles for variously sized tubes, generally ranging from 13 × 75 mm to 16 × 100 mm—sizes that are consistent with the CLSI Standard AUTO01-A. The industrial, motor-driven conveyor belts used in typical LASs have transport rates of a few hundred pucks per hour up to 2000 per hour. In addition, linear synchronous motors and magnetic pucks are now being used in our clinical laboratory ( https://www.aruplab.com/news/06-22-15 ). This technology, which is considerably faster than present conveyors and has virtually no moving parts, may become the next generation of specimen transport for clinical LASs.
Two types of conveyor belt systems are typically used as illustrated in Fig. 29.4 . Fig. 29.4A depicts a loop conveyor that has a single module for both input of new specimens and removal of completed specimens. The flow around the loop takes specimens to the processing and analytical modules. Specimens may be sampled directly by the analytical instrument while on the conveyor, or a robot attached to the workstation may remove selected specimens from the conveyor for analysis. Fig. 29.4B depicts a unidirectional conveyor in which specimens are inputted at one end of the belt, flow past various processing and analytical modules, and arrive at the opposite end where they are removed. Analyzers access specimens in the same way as the loop conveyor. Depending on the vendor, both of these conveyor types can have bypass lanes that enable specimens not designated to stop at certain modules to bypass them entirely and proceed to other modules. This approach has the advantage that it does not require that specimens be aliquoted because specimens pass by all workstations where tests are performed. If specimens are robotically removed from their carriers on the line for testing, systems for queuing empty carriers to return the tubes back to the conveyor and to identify which specimen is in each carrier are required.
Transfer of specimens from the conveyor belt to the laboratory workstation has been implemented in various ways. For example, some manufacturers have equipped their laboratory instruments with devices to move specimen containers from conveyor belt systems onto the analyzers, and others have installed the ability for the analyzer to sample directly from specimen containers remaining on the conveyor. In practice, the automation system requires a device that stops the tube in the exact location required by the analyzer for sample transfer and verifies and transfers the tube’s bar code identification to the analyzer. The CLSI Standard AUTO05-A specifies a common “point in space” that is this exact location for all analyzers and automation conveyor systems.
Depending on the particular automation system, specimens for “STAT” testing may or may not be processed and analyzed at speeds that are faster than if manually handled. Laboratories that are purchasing LASs are advised to learn how each system being considered handles STAT specimens.
The automated capability to store and retrieve specimens on demand and with readily known exact storage locations is an important aspect of automated specimen delivery systems. In addition to automated storage and retrieval options in some of the integrated systems described earlier, several automated or semiautomated options for storage and retrieval, as well as LIS modules and PC-based software systems, permit laboratories to track trays or racks of specimens in their own freezers or refrigerators. Some large reference laboratories have adapted large storage systems commonly used in other industries into their laboratory settings. One such large system, in use in the authors’ laboratory (ARUP), is depicted in Fig. 29.5 .
In this section, the practical considerations that influence a laboratory’s decision to automate part or all of its operations are discussed. A 2007 review outlined how a large hospital laboratory evaluated its needs for automation; the selection and decision processes used by that hospital provide a useful supplement to this discussion.
Any consideration of total or modular laboratory automation should start with an evaluation of requirements. Such an evaluation begins with mapping of the current laboratory workflow from the arrival of patient specimens through completion of testing and reporting of results. Box 29.6 lists potential workflow steps that should be mapped. Mapping of material (specimen) flows and data flows is directly related to process flow and assists the laboratory in determining process steps that (1) are bottlenecks, (2) waste labor, and (3) are prone to error. Workflow mapping thus enables the laboratory to better identify what steps should be considered for automation.
Unpacking from transport containers
Presorting
Temperature preservation
Order entry
Document management (e.g., requisitions)
Labeling
Sorting
Centrifugation
Labeling of aliquot tubes
Decapping
Pouring of aliquots
More sorting
Delivery to laboratory sections
More sorting
Preparing work lists
Labeling analyzer-specific tubes for specimens
Pouring or pipetting analyzer-specific specimens
Loading tubes on analyzers
Performing tests (steps such as extraction, centrifugation, precipitation, and dilution are not specifically listed)
Unloading analyzers
Recapping
Data manipulations (calculations)
Results review and verification
Reporting of results
Delivery of specimens to archival storage system
Archival storage of specimens
Reflexive testing
Repeat testing, diluting, if necessary
Additional physician-ordered testing
Specimen retrieval for additional or repeat testing
Disposal of expired specimens
Some laboratorians use an “80% rule” in guiding decisions about automation. Clinical laboratories have many exceptional tests, specimen containers, and handling situations. Nevertheless, if 80% of the daily workload of specimen containers, covering most routine handling situations, can be standardized and automated, the laboratory will achieve a dramatic reduction in its labor usage and costs, which should be sufficient to justify the investment in automation and the planning and evaluation time involved.
After the laboratory’s existing workflow has been mapped and its requirements have been identified, alternative solutions are considered. Consideration of lean processing and efficiency can help to ensure that automated solutions ultimately improve operations. Vendors are invited to make presentations and to host visits of the laboratory management team to other laboratories where vendors have successful installations. It is important at this stage to focus on the requirements identified by workflow mapping and to not allow the vendor to try to sell equipment that may not be necessary. Key points to remember when a laboratory is considering either new automation or an addition to existing automation are listed below. Although not every circumstance may require assessment of each of these points, the list is a useful outline of a systematic approach to automation that has worked well in the authors’ laboratory and others.
When considering new automation or expansion of existing automation:
Understand the laboratory’s needs (moving from the current state to the desired state and plans for future growth and/or expansion).
Review the laboratory’s specimen volumes by hour and by day (determine peaks and troughs).
Review the percentages that are centrifuged, aliquoted, refrigerated or frozen, and shared between lab sections.
Consider logistics, handling, facilities, and space issues.
Map the laboratory’s workflow. (What are the paths for all specimens?)
Time the workflow to find bottlenecks and time wasters.
Rank needs in order of importance and differentiate needs that are “must haves” from needs that are “desired.”
Identify possible solutions that will meet the identified needs.
Evaluate alternatives that will meet those needs.
Use performance measures for productivity, turn-around time, and quality to establish baselines against which post-automation performance can be measured.
Cost justification will likely be required by management, but the performance measures, if credibly predicted, will likely provide the basis for justification.
Building a highly integrated laboratory generates many potential problems. Most vendors of clinical LASs prefer customer settings in which the integrated analyzers are their own brand. However, many laboratories may prefer to use analyzers from multiple vendors, including the automation system vendor, making integration of instruments and robotic devices from different manufacturers necessary. Decisions must be made concerning which device will be the master controller and which vendor will develop the software that provides overall control of the automation scheme. In addition, individuals or firms that will be responsible for configuration of the automation to the geometry and production schedule of the laboratory must be recruited and trained. Over the past decade, automation vendors have gained considerable experience implementing integrated systems. Although some clinical laboratories have had unsatisfactory experiences, integrated systems are becoming more commonplace.
Readers are referred to the CLSI standard AUTO03-A2 and, in particular, to the Functional Control Model (Section 5.2), which describes the relationships among the LIS, LAS, and various devices. Most often, the LAS has the requisite process control software to support automation. The functional control model, which is depicted in Fig. 29.6 , supports analytical instruments that may be physically attached to the automation system and analyzers that may not be attached but are still interfaced to the LIS. The model does not give dominance to the LIS or the LAS but rather allows for essential information flow in either direction to make the most efficient use of the strengths of each system. Although not specified in the AUTO03-A2 standard, some vendors have implemented redundant clustered computers as their LAS to protect customers from downtime associated with failure of a single computer.
One objective in developing an integrated laboratory is to link laboratory instruments and devices into an automated system to maximize the number of functions automated. Automatic specimen introduction requires the development of mechanical interfaces between each laboratory analyzer and devices such as conveyor belts, mobile robots, or robot arms. Some systems have added enhancements such as electronic interfaces for laboratory instruments to allow remote computer control of front-panel functions, notification of instrument status information, and coordination of the distribution of specimens between instruments. In the ideal integrated clinical LAS, the LAS may be a process controller that would integrate, automate, and monitor many of the decision-making tasks that occur in the daily activity of a laboratory. For example, the LAS would notify an operator that a particular analyzer (even if it is from a different vendor) may be running low on a reagent or may have a fault condition. The LAS (process controller) would control and schedule nonanalytical modules such as automated centrifuges, aliquotters, decappers, and so forth. Most existing LISs have no process control capability, and their interfaces with laboratory analyzers provide only the ability to download accession numbers and the tests requested on each specimen and to upload results generated by the analyzer. The distribution of tasks must be carefully specified in developing such a communications network.
Automation of analytical processes through instrumentation is now standard in most clinical laboratories that perform diagnostic testing. The following sections describe basic concepts inherent to automating the analytical process, with a focus on how these concepts have been integrated into clinical chemistry and immunochemistry instrumentation and automated solutions. Advances in analytical automation in clinical chemistry, immunochemistry and immunology, urinalysis, hematology, coagulation, transfusion medicine, microbiology, mass spectrometry (MS), and molecular diagnostics are then discussed separately.
Become a Clinical Tree membership for Full access and enjoy Unlimited articles
If you are a member. Log in here