Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Artifacts are common in ultrasound imaging.
Knowledge of artifacts can aid in diagnosis.
Understanding ultrasound physics allows for correction of artifacts that distort the visualized anatomy.
ASSUMPTIONS IN GRAY-SCALE IMAGING
Velocity of Sound
Attenuation of Sound
Path of Sound
Beam Profile
PROPAGATION VELOCITY ARTIFACT
ATTENUATION ERRORS
Shadowing
Increased Through Transmission
PATH OF SOUND-RELATED ARTIFACTS
Mirror Image Artifact
Comet-Tail Artifact
Refraction
Anisotropy
Reverberation Artifact
GAS-RELATED ARTIFACTS
Reverberation Artifact
Ring-Down Artifact
Dirty Shadowing Artifact
BEAM PROFILE–RELATED ARTIFACTS
Side Lobe and Grating Lobe Artifacts
Partial Volume Averaging
DOPPLER IMAGING ARTIFACTS
Loss or Distortion of Doppler Information
Artifactual Vascular Flow
Tissue Vibration Artifact
Aliasing and Velocity Scale Errors
Spectral Broadening
Blooming Artifact
Twinkle Artifact
Acknowledgment
Almost all ultrasound images contain artifacts. Many of these artifacts go unrecognized because they are contained within (and contribute to) the background noise. However, when artifacts substantially alter the signal, they become recognizable on images. Certain artifacts distort the images and must be recognized in order to improve image quality or prevent a false diagnosis. Other artifacts add useful data and thus are important for understanding the composition, anatomy, and pathology of the image being visualized. We hope you enjoy this virtual chapter, which encompasses animated illustrations of how artifacts are produced as well as multiple examples of images and videos from the text.
Several assumptions about the propagation of sound waves are made when ultrasound equipment maps echoes onto the image. When one or more of these assumptions prove invalid, artifacts result.
The speed of sound throughout the tissues is assumed to be uniform at 1540 m/sec. When the waves travel through tissues that substantially alter sound velocity, propagation velocity artifacts occur.
Sound waves normally become fainter—that is, their intensity decreases—as they travel through tissues. The equipment assumes that this attenuation of intensity occurs at a constant rate. If the waves travel through tissues that either do not attenuate as much or attenuate more than the adjacent tissues, attenuation errors such as increased through transmission or shadowing arise.
In creating the image, the equipment assumes that the generated sound wave travels from the surface of the probe in a straight line, is reflected off a reflector only once, and returns directly back to the probe at the same angle exactly to the point from which it left the probe. If the sound waves undergo more than one reflection, then artifacts such as mirror image, reverberation, or comet-tail occur. And if the direction of the beam or its echo is altered, refraction or anisotropy artifact may be produced.
An assumption is made that the sound beam generated by the transducer is a narrow line. When the beam is not sufficiency narrow in the imaging plane or in the elevation plane (i.e., along the short axis of the transducer), side lobe or grating lobe or partial volume averaging artifacts may occur.
Ultrasound image processing assumes that the speed of sound in tissue is a constant 1540 m/sec. However, when a sufficiently large structure composed of tissue that propagates speed at a different velocity is encountered, then propagation velocity artifact can occur. In the case of a fatty lesion, the slower speed of propagation (approximately 1450 m/sec) means that the echo will take longer to return to the transducer; thus the lesion is displayed deeper in the image than its true location. Some newer ultrasound machines with multibeam (spatial compounding) features and improved signal processing can minimize this artifact.
Click here to see an explanatory video of propagation velocity artifact ( ).
When the ultrasound beam encounters a focal lesion that attenuates the sound to a greater or lesser extent than the surrounding tissues, the intensity of the beam deep to the lesion will be either weaker (shadowing) or stronger (increased through transmission) than in the surrounding tissues.
Click here to see an explanatory video of attenuation-related artifacts ( ).
Shadowing results when there is a marked reduction in the intensity of the ultrasound deep to a strong reflector, attenuator, or refractor. Clean dark shadows will be seen behind calcified objects when the focal zone is at or just below the structure.
Click here to see an explanatory video of shadowing ( ).
Click here to see real-time cine clip of showing in region of cauda equina ( ).
Edge shadowing is caused by excessive refraction and commonly occurs from the edges vessels, cystic structures, and bones.
Increased through transmission occurs when an object (such as a cyst) attenuates the sound waves less than the surrounding tissues.
Click here to an explanatory video of increased through transmission ( ).
Become a Clinical Tree membership for Full access and enjoy Unlimited articles
If you are a member. Log in here