Intravascular Ultrasound for Imaging of Diseased Arteries


Intravascular ultrasound (IVUS) imaging continues to develop and provides an important perspective of cardiovascular disease. High-resolution transmural images of the vessel allow real-time investigation of pathology and the effects of interventions. The thrust of current IVUS development is to incorporate it as an adjunct to coronary, aortic, and peripheral procedures, providing vessel wall morphometry, endovascular device guidance, and accurate assessment of the intervention.

Intravascular Ultrasound Imaging

In the early 1970s, 360-degree cross-sectional intraluminal ultrasound imaging of vascular structures became available. Cross-sectional imaging was achieved by scanning the ultrasound beam through a full circle and synchronizing the beam direction and deflection on the display, either by mechanically rotating the imaging elements or by electronically switching a transducer array.

Electronically switched phase array devices use frequencies of 8.2 to 40 MHz in 3.5- to 9-Fr catheters to produce cross-sectional images of a vascular segment. The catheters are flexible and track over appropriate-size guidewires. Mechanical transducers use a small transducer that is rotated at the tip of the catheter using a flexible, high-torque cable that extends the length of the device; these devices track over appropriate-size guidewires, but in some devices, the guidewire can create an image artifact.

Several studies have reported that IVUS is accurate in determining the luminal and vessel wall morphology of normal or minimally diseased arteries both in vitro and in vivo. In muscular arteries, distinct sonographic layers are visible, with the media appearing as an echolucent layer between the more echodense intima and adventitia. Even small intimal lesions such as flaps or tears are well visualized because of their high fibrous-tissue content and the difference in echoic properties of these structures when compared with surrounding blood. This three-layer appearance is not readily seen in larger vessels, such as the aorta, because of the increased elastin content in the media ( Figure 1 ).

FIGURE 1, A, Intravascular ultrasound image demonstrating three-layer appearance of iliac artery. The media (M) can be observed as the dark ring surrounding the minimally diseased artery. Areas of soft intimal hyperplasia (H) can be seen at 3 and 9 o’clock positions. The double arrow represents vessel lumen. B, The same vessel at a different location demonstrating hard calcified lesion (C) at 9 o’clock.

Intravascular ultrasound devices are sensitive in differentiating calcified and noncalcified vascular lesions. Because the ultrasound energy is strongly reflected by calcific plaque, it appears as a bright image with dense acoustic shadowing behind it. Gussenhoven and colleagues described four plaque components that can be distinguished using 40 MHz IVUS in vitro: echolucent, which is lipid; soft echoes, fibromuscular tissue, and intimal proliferation, including varying amounts of lipid; bright echoes, collagen-rich fibrous tissue; and bright echoes with acoustic shadowing, calcified tissue (see Figure 1 ).

Advances in image processing and computer technology have resulted in the availability of longitudinal gray-scale IVUS image construction. Longitudinal reconstructions of the IVUS images are obtained by slowly withdrawing the catheter through a vessel. A computer acquires the images, rotates the images 90 degrees, and stacks them to display a longitudinal gray-scale reconstruction. If the pull through is performed with an integrated pullback sled, distances can be measured on the reconstructed image. The current pullback sleds operate at speeds too slow to be useful for peripheral or aortic procedures. For these cases, the investigator can obtain a longitudinal image by pulling the catheter through the vessel by hand, but distances between points cannot be measured. Longitudinal gray-scale IVUS has a unique potential to display vascular pathology, distribution, and tissue characteristics along the length of the vessel, which can be visualized much like an angiogram but with associated wall morphometry ( Figure 2 ).

FIGURE 2, A, Angiogram of abdominal aortic aneurysm. B, Longitudinal gray-scale reconstruction of intravascular ultrasound images along the length of an abdominal aortic aneurysm demonstrating renal artery, aneurysm, iliac bifurcation, and left hypogastric artery. C, Renal vein on top of the aorta and left renal artery orifice. D1, Location just below renal artery for diameter measurement. D2, Diameter measurement 1.5 cm below renal artery. Note measurement in short axis as a result of angulations, similar to computed tomography measurements. E, Aneurysm; note absence of wall from 1 to 5 o’clock positions. F, First ring of left iliac artery. G, Left hypogastric artery orifice.

Virtual histology is an imaging option available with the phase-array IVUS catheters. It uses spectral analysis of the radiofrequency ultrasound backscatter signals to identify components of atherosclerotic lesions. The system color codes the four different tissue components: fibrous (green), fibrofatty (light green), dense calcium (white), and necrotic core (red) ( Figure 3 ). The color-coded map helps the investigator understand how the lesion will react to the selected treatment: Will the lesion resist stent deployment or will it be liable to embolization? It is postulated that acute coronary syndromes may be caused by the lipid-rich necrotic core and a rupture of the overlying fibrous cap following angioplasty. This type of plaque differentiation using virtual histology might also play a future role in the selection of carotid stent procedures as well as peripheral vascular interventions (see Figure 3 ).

FIGURE 3, A, Angiogram of superficial femoral artery prior to treatment. B, Angiogram following six passes of the atherectomy device demonstrating a 50% residual stenosis (arrow) (Turbo Hawk, ev3 Endovascular, Plymouth, MN). C, Corresponding intravascular ultrasound (IVUS) demonstrating 70% residual luminal stenosis. D, The virtual histology (VH) image of the same lesion demonstrating areas of fibrous (green), fibrofatty (light green), necrotic core (red), and dense calcium (white). E, Angiogram following four additional passes of the device, which resulted in a similar angiographic image. F and G, IVUS and VH images demonstrating increase in luminal area with residual 52% stenosis (at arrow in image E).

Clinical Utility

Intravascular ultrasound imaging’s advantages over other vascular imaging modalities include defining the transmural distribution of disease within the vessel, characterizing plaques and intimal lesions, and providing accurate cross-sectional information regarding luminal and vessel wall morphology before and after endovascular interventions.

Contrast arteriography, the standard for imaging the distribution and severity of vascular lesions, is accurate in defining vessel luminal dimensions if the luminal profile is circular. In advanced disease, however, it provides limited information regarding the morphology and extent of disease in the arterial wall, aside from documenting visible calcification and the topography of the luminal surface. It has been shown that arteriography is restricted in imaging vessels with substantial luminal ellipticity, usually overestimating the luminal cross-sectional areas and underestimating the degrees of stenosis.