Cardiac CT (CCT) scanning remains problematic in atrial fibrillation (AF) in several ways because of inadequate temporal resolution:
The heart rate tends to be higher in patients with AF ( Table 22-1 ).
ATRIAL FIBRILLATION ( n = 24) | SINUS RHYTHM ( n = 119) | P | |
---|---|---|---|
Pre-scan heart rate | 54 ±14 bpm | 62 ±112 bpm | <.01 |
Postinjection heart rate | 103 ± 30 bpm | 71±17 bpm | <.01 |
Nondiagnostic CTA studies | 8% | 10% | ns |
Good or excellent CTA image quality | 54% | 79% | .01 |
Radiation exposure | 26 ± 8 mSv | 8 ± 4 mSv | <.01 |
The cardiac cycle length is always irregular.
The increment in heart rate is greater during contrast injection and during image acquisition.
Opacification of the left atrial appendage (LAA), which usually is dilated and nearly effectively stagnant, is delayed well beyond that of the left heart cavities and coronary arteries, potentially resulting in pseudothrombus in the left atrium, the result of incomplete opacification of the LAA due to enlargement of the LAA and altered flow characteristics in the left atrium due to AF ( Table 22-2 ).
ATRIAL FIBRILLATION | SINUS RHYTHM | ||
---|---|---|---|
Left atrial appendage volume | 15 ± 7 mL | 7 ± 3 mL | 0.02 |
(Proven) pseudothrombus in the left atrial appendage during CTA without delayed imaging | 71% | 0 |
With different cardiac cycle lengths, multicycle/segmented acquisition image quality is seriously compromised. Using 64-slice CCT, overall only half (54%) of coronary CT angiography (CTA) image studies of AF are of adequate or good quality. Use of retrospective imaging at full tube output increases radiation exposure. Although feasible, cardiac/coronary imaging during AF with CT scanning requires further development to obtain increased yield and to lessen radiation exposure.
Structural noncoronary studies such as pulmonary venous and coronary venous anatomy studies are more likely than coronary CTA studies to be adequate.
However, using 320-slice CCT, the large majority of cases can be adequately imaged despite the presence of AF ( Table 22-3 ).
AUTHOR | SENSITIVITY (%) | SPECIFICITY (%) | PPV (%) | NPV (%) | EVALUABLE SEGMENTS (No.) | DOSE (mSv) |
---|---|---|---|---|---|---|
Xu et al. | 90 | 99 | 86 | 99 | 97 | 13 ± 5 |
Pasricha et al. | 99 | 19 |
The distal left circumflex artery appears to pose the most problems in evaluating for the presence and severity of coronary artery disease (CAD).
The presence of underlying AF challenges the accurate detection of atrial appendage thrombi. In contrast to patients in sinus rhythm, the variable cardiac cycle length and greater and more variable heart rate increase in response to the stimuli of scanner acceleration and contrast injection of patients in AF excessively tax the ability of 64-slice CT scanners to accurately assess the LAA. The left atrium, whose imaging is less affected by AF than is that of the appendage, is more readily assessed for the presence or absence of thrombus. The stagnation of flow within the LAA necessitates a delayed scan at 2 or more minutes post-injection, with additional radiation exposure.
Stagnation of flow within the LAA is a major confounder of assessment of the LAA for thrombosis. The delayed clearance of non–contrast-enhanced blood within the LAA has established the high incidence of false positive studies without delayed images. For example, Martinez et al. found that in 402 consecutive patients undergoing non–ECG-gated CCT and transesophageal echocardiographic (TEE) pre-pulmonary venous isolation, CCT identified 40 LAA filling defects, of which only 9 were confirmed by TEE. False negatives also are common; for example in Tang et al.’s study, only 4 of 10 thrombi seen on TEE were seen on CCT. Some false-positive defects persist even after delay.
There are few data to this point, but small early studies have shown correlation between TEE and contrast cardiovascular CT (CVCT) measurements of LAA size and correlation of findings of thrombi larger than 7 mm. Higher heart rates and irregular rhythm (the norm in AF) challenge feasibility at this point. ECG gating is important. CVCT is unable to show left atrial or appendage spontaneous contrast.
Small accessory LAAs are recognized by 64-slice CCT, but these, due to their novel recognition, are still of unknown significance. Accessory LAAs have been reported in 10% to 15% of the adult population undergoing cardiac CT ( Figs. 22-1 through 22-4 ).
Atrial fibrillation usually is initiated by ectopic activity within the proximal pulmonary veins. Catheter-based techniques, in which access to the left atrium occurs via a transseptal puncture of the interatrial septum, are able to deliver radiofrequency current (or cryotherapy) to the sites that have been identified by mapping techniques as responsible for AF, electrically isolating them from the remainder of the atria. The usual success rate in achieving sinus rhythm at 1 year with two or three attempts is 75% to 85%—a marked improvement compared to medical therapy, which affords about 20% success.
Anatomic features are not adequate to identify the site responsible for AF.
Conventionally, real-time retrograde injection of contrast dye into the left atrium has been used during the procedure to visually guide catheter placement. However, opacification of the right inferior pulmonary vein is difficult to achieve by reflux venography. Other modalities such as intracardiac echocardiography and TEE are less robust than reflux venography, which is less robust than CCT in detecting normal and variant anatomy (e.g., supernumerary and common ostia and branch variation) ( Table 22-4 ).
IMAGING MODALITY | PERCENTAGE DETECTION OF ALL FOUR PULMONARY VENOUS OSTIA ( n = 24) |
---|---|
Cardiac CT | 98 |
ICE | 93 |
TEE | 81 |
Reflux venography | 71 |
Variations in pulmonary venous anatomy are fairly common. The most common variant is a separate right middle pulmonary vein. Pulmonary venous variants include:
Three to five right-sided ostia (28%)
One or two separate middle lobe pulmonary venous ostia (26%)
A single right ostium on the right side (2%)
A single left-sided ostium (14%)
A right superior supranumerary pulmonary vein (2.2%), which passes behind the bronchus intermedius and drains mainly the posterior segment of the right upper lobe, but also receives a few subsegmental branches from the superior segment of the right lower lobe (mean diameter: 5.1 mm)
Right pulmonary vein to the superior vena cava
Right pulmonary vein to the right atrium
Right pulmonary vein to the inferior vena cava or a hepatic vein (i.e., scimitar syndrome)
Left pulmonary vein to the innominate vein (i.e., vertical vein)
The size (diameter) of pulmonary veins depends on:
Volume status
Respiratory phase (inspiration decreases pulmonary venous return and size). Expiratory phase images provide a better match to MRI depiction of pulmonary veins.
The phase of the cardiac cycle (by 25%–45%):
Maximal size of pulmonary veins: 35%
Minimal size of pulmonary veins: 85%
The use of image guidance during pulmonary venous isolation/ablation procedures reduces fluoroscopy times and increases success.
CCT has become the principal noninvasive diagnostic test to depict venous anatomy of the heart for the following reasons:
Especially when compared with TEE, intracardiac echocardiography, and reflex venography, it is less operator dependent.
Unlike cardiac MR (CMR), it can be performed on patients with pacemakers and implantable cardioverter defibrillators.
Its results are reproducible.
It can depict the thickness of the wall of the pulmonary veins.
It can characterize the lung parenchyma for complications such as pulmonary edema and infarction.
It can image the mediastinum both before and after ablation. This can assist in visualizing the relation of the esophagus to the left atrium and in detecting the presence of large intrathoracic hiatus hernias.
CCT can offer several forms of depiction of pulmonary venous and left atrial anatomy that may be useful to interventional electrophysiologists before they undertake catheter-based ablation procedures and to interventional cardiologists:
Three-dimensional (3D) reconstruction of surface views depicting the coronary venous structures on the myocardium
“Endoscopic” views to depict pulmonary veins from a perspective within the left atrium
3D views of pulmonary veins and the left atrium, with or without the remainder of the heart
Diagnostic imaging of pulmonary vein stenosis, achieved with multiplanar reconstructions, maximal intensity projections, and volume-rendered images
Examples of normal pulmonary veins are shown in Figures 22-5 and 22-6 .
The size of pulmonary veins is variable and is influenced by several factors ( Table 22-5 ).
LARGER | SMALLER |
---|---|
Chronic long-term atrial fibrillation | Chronic short-term atrial fibrillation |
Chronic atrial fibrillation | Paroxysmal atrial fibrillation |
Larger LA | Smaller LA |
Mitral disease | No mitral disease |
LV disease | No LV disease |
Volume loaded | Volume depleted |
Male | Female |
There is a wide range of variant pulmonary venous anatomy, and image examples of variant pulmonary venous anatomy are seen in Figures 22-7 through 22-12 .
Impact of respiration on pulmonary venous morphology:
Inspiration:
Increases the intervein distance (2–4 mm)
Increases the angle between (“splays”) the veins
Decreases the diameter of the veins
Impact of the cardiac cycle on pulmonary venous ostia:
Enlarges at end-systole versus end-diastole (35–45%)
Maximum diameter, on average, at 35% RR interval
Minimum diameter, on average, at 85% of RR interval (atrial contraction post–early diastolic emptying into the LV)
In pre-ablation cases, CCT affords more imaging information than does CMR, albeit at the expense of increased radiation exposure ( Table 22-6 ).
CMR | CTA | |
---|---|---|
Radiation dose (mSv) | 0 | 0.7–6 |
Spatial resolution (mm) | 1.5–3 | 0.5–0.7 |
Gating/motion compensation | None | May be ECG-gated |
Evaluation of lungs | Limited | Excellent, within the obtained FOV |
Evaluation of mediastinum | Limited | Excellent, within the obtained FOV |
Evaluation of coronary artery anatomy | Poor | If gated, especially if 320-slice CCT, excellent |
Evaluation of sinoatrial coronary branch | None | If gated, and heart rate is low, can be imaged |