Vascular and cardiac blood pool contrast enhancement is determined by the interaction of iodine administration and blood flow. The goals of contrast administration are:
Adequate opacification of cavities and vessel lumen
Avoidance of both under- and over-opacification. It is important to recognize that excessive contrast concentration generates high attenuation artifact.
Starburst high-attenuation artifacts from the superior vena cava (SVC) are among the most common of all annoying and frustrating artifacts.
CT angiography (CTA) requires high left heart/aortic contrast and low right heart contrast (high right heart opacification confounds assessment of the right coronary artery and the right ventricular contours).
Optimal enhancement depends on the purposes of the study.
The optimum contrast effect for vascular enhancement is 300 to 500 HU.
The optimal contrast effect for coronary plaque characterization is less: 250 to 350 HU.
Optimal opacification requires a near-steady level of contrast throughout the entire duration of the study.
Synchronization of opacification to image acquisition phase is crucial. For CTA, the principal challenge is to synchronize these three variables:
The arrival of the contrast medium in the left heart
Washout from the right heart with the saline push
Acquisition phase
As scanners are developed that can use larger numbers of slices and provide greater breadth of scanning, acquisition time is falling, as is the needed duration of contrast effect, and, therefore, the volume of contrast needed. In comparison to 10-slice scanners, 64-slice scanners need only half the volume of contrast. In comparison to 16-slice scanners, 64-slice scanners need one third less contrast.
Larger field of view: e.g., post-aortocoronary bypass studies where the field of view along the Z-axis is from the subclavian arteries to the base of the heart (to visualize internal thoracic arteries as well as the heart)
Slower (fewer detectors) scanners
Lesser pitch
Faster (more detectors) scanners
Smaller scan length
Greater pitch
A robust protocol is wiser than an ambitious protocol.
A dual-head injector is preferable (for contrast and for saline).
The greater the kV, the greater the penetration and the lesser the contrast effect.