Recognizing the Correct Placement of Lines and Tubes and Their Potential Complications: Critical Care Radiology


Patients in the critical or intensive care units (ICU) are monitored on a frequent basis with portable chest radiography both to check on the position of their multiple assistive devices and to assess their cardiopulmonary status.

Case Quiz 9 Question

This patient just had a central venous catheter inserted on the left side in the subclavian vein (arrow) and you are asked to check this postinsertion radiograph to see if it is correctly positioned. What is your answer? The correct answer appears at the end of this chapter.

  • Diseases commonly seen in critically ill patients are discussed in other chapters ( Table 9.1 ).

    TABLE 9.1
    Common Diseases in Critically Ill Patients
    Finding or Disease Discussed in
    Acute respiratory distress syndrome Chapter 11
    Aspiration Chapter 8
    Atelectasis Chapter 6
    Congestive heart failure (pulmonary edema) Chapter 11
    Pleural effusion Chapter 7
    Pneumomediastinum Chapter 24

  • In this chapter, you’ll get practical advice for evaluating the successful (or unsuccessful) insertion and ultimate position of multiple tubes, lines, catheters, and other supportive apparatus used in the ICU. Information for even more devices can be found online ( e-Fig. 9.1 to e-Fig. 9.10) .

e-Fig. 9.1
Aortic Valve Replacement.
(A) There is a prosthetic valve overlying the spine and obliquely oriented in the direction of the aortic valve (circle). (B) A line drawn between the carina and the anterior cardiophrenic sulcus (red line) can be helpful in identifying the location of a prosthetic valve; the aortic valve (A) (circle) is usually above the line while the area of the mitral valve (M) is below the line.

e-Fig. 9.2
Tricuspid and Mitral Valve Replacements.
The tricuspid valve is oriented almost vertically on the frontal view (white arrow in A) and anteriorly on the lateral view (black arrow in B). The mitral valve is usually found to the left of the spine on the frontal view (black arrow in A) and posteriorly on the lateral view (white arrow in B). Note the markedly enlarged left atrial appendage (dashed white arrow) in this patient with severe mitral stenosis. There is an epicardial pacemaker lead also seen, identifiable by its “corkscrew” end (circle).

e-Fig. 9.3
Pulmonic Valve Replacement.
The pulmonic valve is more superior to the other cardiac valves. It is seen in (A) in the black circle and (B) in the white circle. The pulmonary valve is not replaced as frequently as the aortic or mitral valves. The most common indications for surgical pulmonary valve replacement are pulmonary regurgitation with or without stenosis, right ventricular dilatation, and arrhythmias. Some prosthetic valves are now being replaced by nonsurgical, catheter-based deployment.

e-Fig. 9.4
Left Ventricular Assist Device (LVAD).
Surgically implanted, mechanical circulatory assist devices in patients with severe and medically refractory heart failure are now used as a permanent alternative to heart transplantation. The device shown here, with the image’s black and white inverted, is the HeartMate. The pump is implanted in the preperitoneal space or peritoneal cavity below the heart (dashed arrow). It is a continuous flow pump with no pulsatile action; therefore the patient who receives such a device will have no palpable pulse. The inflow cannula is anastomosed to the tip of the left ventricle (solid arrow). The outflow cannula is attached to the ascending aorta (dashed arrow). The unit can run on battery or external power.

e-Fig. 9.5
Impella Pump.
This is a catheter-inserted, left ventricular assist pump. The pump can be inserted through the femoral artery, then fed up the descending aorta to the ascending aorta (black arrows), across the aortic valve and into the left ventricle (white arrows) to terminate in the left ventricle (circle). It draws blood across the aortic valve. Like an intra-aortic balloon pump (IABP), it is usually used for temporary circulatory support in patients who develop cardiogenic shock after cardiac surgery and who have not responded to standard medical therapy.

e-Fig. 9.6
Left Atrial Occlusion Clip, Frontal (A) and Lateral (B) Views.
The clip can be used in patients with atrial fibrillation since the left atrial appendage is the source of over 90% of the thrombi produced by that arrhythmia. The clip (circle and arrows) is applied during surgery to the outside surface of the left atrial appendage and is intended to permanently occlude the opening to the appendage at its base, thus preventing blood from entering it.

e-Fig. 9.7
Implantable Loop Recorder ( Circle ).
This is a subcutaneous electrocardiographic monitoring device that stores ECG data automatically in response to specific rhythm abnormalities or in response to the patient manually activating it. It is mainly used for diagnosis in patients with recurrent, unexplained episodes of syncope or palpitations, for long-term monitoring in patients with documented or suspected atrial fibrillation, and for risk identification in patients who have had a myocardial infarction.

e-Fig. 9.8
Coronary Artery Stent, Frontal (A) and Lateral (B) Views.
There is a stent in the left anterior descending coronary artery (arrows). Not all stents will be visible on chest radiographs and many newer bioabsorbable stents will become invisible to all modalities over time. Most newer coronary artery stents are MRI-compatible, but that should be determined before ordering any MRI procedure.

e-Fig. 9.9
Tissue Expander.
Following mastectomy, a temporary, saline-filled tissue expander may be inserted beneath the breast skin and muscle (the port is shown by the arrow) and gradually inflated with additional saline. After several months, the expander is removed and the patient receives either breast reconstruction or the insertion of a permanent breast implant. The expander allows for skin matched in color and texture to the opposite side with minimal scarring.

e-Fig. 9.10
PEG (Percutaneous Endoscopic Gastrostomy) Tube (Intra- and Extraluminal).
PEG tubes are used for the long-term enteral nutrition of patients instead of a gastrostomy. To determine whether a PEG tube is in the correct intraluminal location, a small amount of water-soluble contrast may be injected through the tube, sometimes under fluoroscopic guidance. (A) The retention balloon of the PEG tube (arrow) is inflated in the stomach (S). All of the contrast is intraluminal and exits the stomach into the duodenum (D). (B) An injection of contrast into another PEG tube demonstrates all of the contrast is extraluminal in the peritoneal cavity (white arrows). The contrast outlines the external surface of air-filled small bowel (dashed black arrow).

  • Almost always, a conventional radiograph is obtained after the insertion or attempted insertion of one of these devices to check on its position and to rule out any unintended consequences.

  • Therefore for each tube or device, you will learn :

    • Why they are used

    • Where they belong when properly placed

    • How such devices can be malpositioned and what complications may occur from the device ( Box 9.1 ).

      BOX 9.1
      Postprocedure Warning Signs

      • There are many complications besides malpositioning that can follow an invasive procedure, but a few are more common and should raise suspicion when they occur immediately after the procedure.

      • For any tube, line, or device that either enters or passes through the thorax, be alert for postprocedure:

        • Appearance or increase in pleural effusion

        • Pneumothorax

        • Pneumomediastinum

        • Widening of the mediastinal shadow

      • For any tube, line, or device inserted in the abdomen, be alert for postprocedure:

        • Pneumoperitoneum

Endotracheal and Tracheostomy Tubes

Endotracheal Tubes (ETT)

  • Why they are used

    • Assist ventilation

    • Isolate the trachea to permit control of the airway

    • Prevent gastric distension

    • Provide a direct route for suctioning

    • Administer medications

  • Correct placement of an ETT

    • Endotracheal tubes are usually wide-bore tubes (about 1 cm) with a radio-opaque marker stripe and no side-holes. The tip is frequently diagonally shaped.

    • With the patient’s head in the neutral position (i.e., bottom of mandible is at the level of C5-C6), the tip of the ETT should be about 3 to 5 cm from the carina or roughly half the distance between the medial ends of clavicles and the carina ( Fig. 9.1 ).

      Fig. 9.1, Endotracheal Tube in Satisfactory Position.

  • Ideally the diameter of the endotracheal tube should be one-half to two-thirds the width of trachea. An inflated cuff (balloon), if present, may fill—but shouldn’t distend—the lumen of the trachea ( Fig. 9.2 ).

    Fig. 9.2, Endotracheal Tube with Cuff Overinflated.

Important Points

  • How to find the location of the carina on a frontal chest radiograph

    • Follow the right or left main bronchus backward until either meets the opposite main bronchus.

    • Alternatively, the carina projects over the T5, T6, or T7 vertebral bodies in 95% of people.

  • Movement of tip with flexion and extension

    • Neck flexion may cause 2 cm of descent of the tube tip. This is why the tip should be 3 to 5 cm above the carina.

    • Neck extension from neutral may cause 2 cm of ascent of the tip.

    • There is a silly, but helpful, rhyming sentence to help remember the direction of movement of the tip of an ETT with movement of the head: “The tip of the hose (i.e., ETT) follows the tip of the nose.

  • Incorrect positioning and complications of an ETT

    • Most common malposition: because of the shallower angle and wider diameter of the right main bronchus or bronchus intermedius, the tip of the ETT will tend to slide into the right-sided bronchial tree preferentially to the left.

      • This can lead to atelectasis (especially of the nonaerated right upper lobe and left lung) (see Case Quiz 6).

      • Intubation of the right main bronchus could also lead to a right-sided tension pneumothorax .

    • Inadvertent esophageal intubation will produce a dilated stomach.

    • The tip of the tube should not be positioned in the larynx or pharynx as damage to the vocal cords can occur ( Fig. 9.3 ).

      Fig. 9.3, Endotracheal Tube Too High.

Tracheostomy Tubes

  • Why they are used

    • In patients with airway obstruction at or above level of larynx

    • In respiratory failure requiring long-term intubation (>21 days)

    • For airway obstruction during sleep apnea

    • When there is paralysis of the muscles that affect swallowing or respiration

  • Correct placement of a tracheostomy tube ( Fig. 9.4 )

    • The tip should be about halfway between the stoma in which the tracheostomy tube was inserted and the carina. The carina is usually around the level of T3.

    • Unlike an ETT, the placement of the tip of a tracheostomy tube is not affected by flexion and extension of the head.

    • The width of the tracheostomy tube should be about two-thirds the width of trachea.

    Fig. 9.4, Tracheostomy Tube in Correct Position.

  • Incorrect placement and complications of a tracheostomy tube

    • Immediately after insertion look for signs of inadvertent perforation of the trachea such as pneumomediastinum, pneumothorax, and subcutaneous emphysema.

    • If the tracheostomy tube is equipped with a cuff , the cuff should generally be inflated to a diameter that fills , but does not distend, the normal tracheal contour.

    • Long-term complication of tracheostomies:

      • Tracheal stenosis is the most common late-occurring complication of a tracheostomy tube and can occur at the entrance stoma, level of the cuff, or at the tip of tube, but is most common at the stoma .

Intravascular Catheters

Central Venous (Pressure) Catheters (CVC, CVP)

  • Why they are used

    • For venous access to instill chemotherapeutic and hyperosmolar agents not suitable for peripheral venous administration

    • Measurement of central venous pressure

    • To maintain and monitor intravascular blood volume

  • Correct placement of central venous catheters

    • Central venous catheters have a small diameter (3 mm) and are uniformly opaque without a marker stripe.

    • Their correct placement is shown in Fig. 9.5 and Fig. 9.6 .

      Fig. 9.5, Central Venous Catheter (CVC).

      Fig. 9.6, Subclavian Central Venous Catheters in Correct Position, (A) Inserted on Right Side and (B) Inserted on Left.

  • Incorrect placement and complications of central venous catheters

    • Central venous catheters are most often malpositioned with their tips in the right atrium or internal jugular vein ( Fig. 9.7 ). In the right atrium, they can produce cardiac arrhythmias. When central venous catheters are malpositioned, they may provide inaccurate central venous pressure readings.

      Fig. 9.7, Central Venous Catheter Malpositioned in Internal Jugular Vein.

    • Pneumothorax can occur in up to 5% of CVC insertions, more often with the subclavian approach than the internal jugular route.

    • All bends in the catheter should be smooth curves. Occasionally CVCs may perforate the vein and lie outside of the blood vessel. Look for sharp bends/kinks in the catheter as a clue to a potential perforation.

    • Sometimes, they may be inadvertently inserted in the subclavian artery rather than the subclavian vein. Suspect arterial placement if the blood return is pulsatile upon placement and the course of the catheter follows the aortic arch or fails to descend to the right of the spine ( Fig. 9.8 ).

      Fig. 9.8, Arterial Placement of Central Venous Catheter.

Important Points

  • Two or more attempts at inserting a CVC

    • A frontal chest radiograph is obtained following placement of a CVC. Should initial placement fail , it is customary to obtain a chest radiograph before trying insertion on the other side to avoid the possibility of producing bilateral pneumothoraces.

You're Reading a Preview

Become a Clinical Tree membership for Full access and enjoy Unlimited articles

Become membership

If you are a member. Log in here