Digital Image Processing in Nuclear Medicine


Image processing refers to a variety of techniques that are used to maximize the information yield from a picture. In nuclear medicine, computer-based image-processing techniques are especially flexible and powerful. In addition to performing basic image manipulations for edge sharpening, contrast enhancement, and so forth, computer-based techniques have a variety of other uses that are essential for modern nuclear medicine. Examples are the processing of raw data for tomographic image reconstruction in single photon emission computed tomography (SPECT) and positron emission tomography (PET) (see Chapter 16, Chapter 17, Chapter 18 ), and correcting for imaging system artifacts (e.g., Chapter 14 , Section B, and Chapter 18 , Section D). Another important example is time analysis of sequentially acquired images, such as is done for extracting kinetic data for tracer kinetic models (see Chapter 21 ). Computer-based image displays also allow three-dimensional (3-D) images acquired in SPECT and PET to be viewed from different angles and permit one to fuse nuclear medicine images with images acquired with other modalities, such as computed tomography (CT) and magnetic resonance imaging (MRI) (see Chapter 19 ). Computer-based acquisition and processing also permit the raw data and processed image data to be stored digitally (e.g., on computer disks) for later analysis and display.

All of these tasks are performed on silicon-based processor chips, generically called microprocessors. The central processing unit (CPU) of a general purpose computer, such as a personal computer, is called a general purpose microprocessor. Such devices can be programmed to perform a wide variety of tasks, but they are relatively large and not very energy efficient. For very specific tasks, an application-specific integrated circuit often is used. ASICs are compact and energy efficient, but their functionality is hardwired into their design and cannot be changed. Examples of their uses include digitizing signals (analog-to-digital converters) and comparing signal amplitudes (pulse-height analyzers and multichannel analyzers). Other categories of microprocessors include digital signal processors (DSPs) and graphics processing units. These devices have limited programmability, but they are capable of very fast real-time signal and image processing, such as 3-D image rotation and similar types of image manipulations.

The technology of microprocessors and computers is undergoing continuous and rapid evolution and improvement, such that a “state-of-the-art” description rarely is valid for more than a year or, in some cases, even a few months. However, the end result is that the usage of computers and microprocessors in nuclear medicine is ubiquitous. They are used not only for acquisition, reconstruction, processing, and display of image data but also for administrative applications such as scheduling, report generation, and monitoring of quality control protocols.

In this chapter, we describe general concepts of digital image processing for nuclear medicine imaging. Additional discussions of specific applications are found in Chapter 13, Chapter 14, Chapter 15, Chapter 16, Chapter 17, Chapter 18, Chapter 19, Chapter 21 .

Digital Images

Basic Characteristics and Terminology

For many years, nuclear medicine images were produced directly on film, by exposing the film to a light source that produced flashes of light when radiations were detected by the imaging instrument. As with ordinary photographs, the image was recorded with a virtually continuous range of brightness levels and x-y locations on the film. Such images sometimes are referred to as analog images. Very little could be done in the way of “image processing” after the image was recorded.

Virtually all modern nuclear medicine images are recorded as digital images. This is required for computerized image processing. A digital image is one in which events are localized (or “binned”) within a grid comprising a finite number of discrete (usually square) picture elements, or pixels ( Fig. 20-1 ). Each pixel has a digital (nonfractional) location or address, for example, “ x = 5, y = 6.” For a gamma camera image, the area of the detector is divided into the desired number of pixels ( Fig. 20-2 ). For example, a camera with a field-of-view of 40 cm × 40 cm might be divided into a 128 × 128 grid of pixels, with each pixel therefore measuring 0.3125 mm × 0.3125 mm. Each pixel corresponds to a range of possible physical locations within the image. If an event were determined to have interacted at a location x = 4.8 cm, y = 12.4 cm, the appropriate pixel location for this event would be


x-pixel location = 4.8 cm / 0.3125 cm / pixel = int ( 15.36 ) = 15

y-pixel location = 12.4 cm / 0.3125 cm / pixel = int ( 39.68 ) = 40

FIGURE 20-1, A digital image consists of a grid or matrix of pixels, each of size L × L units. Each pixel has an x-y address location, with pixel value, p ( x,y ), corresponding to the number of counts or other quantity associated with that pixel.

FIGURE 20-2, Subdivision of the gamma camera detector area for generating a digital image. The photomultiplier tube signals are analyzed using analog-to-digital converters to assign the digital matrix location for each detected event.

where int( x ) denotes the nearest integer of x , and the pixels are labeled from 0-127 with the coordinate system defined as shown in Figure 20-2 .

A similar format is used for digital multislice tomographic images, except that the discrete elements of the image would correspond to discrete 3-D volumes of tissue within a cross-sectional image. The volume is given by the product of the x- and y-pixel dimensions multiplied by the slice thickness. Thus they are more appropriately called volume elements, or voxels. However, when discussing an individual tomographic slice, the term pixel still is commonly used. In tomographic images, the “intensity” of each voxel may or may not have a discrete integer value. For example, voxel values for a reconstructed image will generally have noninteger values corresponding to the calculated concentration of radionuclide within the voxel.

Depending on the mode of acquisition (discussed in Section A.4 ), either the x-y address of the pixel in which each event occurs, or the pixel value, p ( x , y ), is stored in computer memory. For 3-D imaging modes, such as 3-D SPECT or PET, individual events are localized within a 3-D matrix of voxels, and the reconstructed value in a voxel is denoted as v ( x , y , z ). Depending on how data are acquired and processed by the imaging system, the pixel or voxel value may correspond to the number of counts, counts per unit time, the reconstructed pixel or voxel value, or absolute radionuclide concentrations (kBq/cc or µCi/cc).

Although most interactions between the user and a computer system involve conventional decimal numbers, the internal operations of the computer usually are performed using binary numbers. Binary number representation uses powers of 2, whereas the commonly used decimal number system uses powers of 10. For example, in decimal representation, the number 13 means [(1 × 10 1 ) + (3 × 10 0 )]. In the binary number system, the same number is represented as 1101, meaning [(1 × 2 3 ) + (1 × 2 2 ) + (0 × 2 1 ) + (1 × 2 0 )], or (8 + 4 + 0 + 1) = 13. Each digit in the binary number representation is called a bit (an abbreviation for “ bi nary digi t ”). In general, an n -bit binary number can represent decimal numbers with values between zero and (2 n − 1).

Binary numbers are employed in computer systems because they can be represented conveniently by electronic components that can exist only in an “on” or “off” state. Thus an n -bit binary number can be represented by the “on” or “off” state of a sequence of n such components. To communicate sensibly with the outside world, the binary numbers used within the computer must be converted into decimal integers or into decimal numbers and fractions. The latter are called floating point numbers. The methods by which binary numbers are converted to decimal format are beyond the scope of this presentation and can be found in more advanced texts on computer systems.

Digital images are characterized by matrix size and pixel depth. Matrix size refers to the number of discrete picture elements in the matrix. This in turn affects the degree of spatial detail that can be presented, with larger matrices generally providing more detail. Matrix sizes used for nuclear medicine images typically range from (64 × 64) to (512 × 512) pixels. Matrix size virtually always involves a power of 2 (2 6 and 2 9 in the previous examples) because of the underlying binary number system used in the computer.

Pixel depth refers to the maximum number of events that can be recorded per pixel. Most systems have pixel depths ranging from 8 bits (2 8 = 256; counts range from 0 to 255) to 16 bits (2 16 = 65,536; counts range from 0 to 65,535). Note again that these values are related to the underlying binary number system used in the computer. When the number of events recorded in a pixel exceeds the allowed pixel depth, the count for that pixel is reset to 0 and starts over, which can lead to erroneous results and image artifacts.

Pixel depth also affects the number of gray shades (or color levels) that can be represented within the displayed image. In most computer systems in use in nuclear medicine, 8 bits equals a byte of memory and 16 bits equals a word of memory. The pixel depth, therefore, frequently is described as “byte” mode or “word” mode. *

* Most modern computer CPUs have 32-bit or 64-bit processors. This means they can process data 32 or 64 bits at a time; however, this is largely independent of image display and how pixel values are stored.

Spatial Resolution and Matrix Size

The spatial resolution of a digital image is governed by two factors: (1) the resolution of the imaging device itself (such as detector or collimator resolution) and (2) the size of the pixels used to represent the digitized image. For a fixed field-of-view, the larger the number of pixels, that is, the larger the matrix size, the smaller the pixel size ( Fig. 20-3 ). Clearly, a smaller pixel size can display more image detail, but beyond a certain point there is no further improvement because of resolution limitations of the imaging device itself. A question of practical importance is, At what point does this occur? That is, how many pixels are needed to ensure that significant detail is not lost in the digitization process?

FIGURE 20-3, Digital images of the liver and spleen (posterior view) displayed with different matrix sizes. The larger the matrix size, the smaller the pixels and the more detail that is visible in the image.

The situation is entirely analogous to that presented in Chapter 16 for sampling requirements in reconstruction tomography. In particular, Equation 16-13 applies—that is, the linear sampling distance, d , or pixel size, must be smaller than or equal to the inverse of twice the maximum spatial frequency, k max , that is present in the image:


d = 1 / ( 2 × k max )

This requirement derives directly from the sampling theorem discussed in Appendix F , Section C.

Once this sampling requirement is met, increasing the matrix size does not improve spatial resolution, although it may produce a cosmetically more appealing image with less evident grid structure. If the sampling requirements are not met (too coarse a grid), spatial resolution is lost. The maximum spatial frequency that is present in an image depends primarily on the spatial resolution of the imaging device. If the resolution of the device is specified in terms of the full width at half maximum (FWHM) of its line-spread function ( Chapter 15 , Section B.2), then the sampling distance (pixel size) should not exceed about one third of this value to avoid significant loss of spatial resolution , that is,


d FWHM 3

This applies for noise-free image data. With added noise it may be preferable to relax the sampling requirement somewhat (i.e., use larger pixels) to diminish the visibility of noise in the final digitized image.

Example 20-1

What is the approximate spatial resolution that can be supported for a 30-cm diameter field-of-view using a 64 × 64 matrix? A 128 × 128 matrix? Assume that the original data are noise free.

Answer

64 × 64 matrix

A 64 × 64 image matrix results in a pixel size of 300 mm/64 = 4.69 mm. From Equation 20-2 , this would be suitable for image resolution given by


FWHM 3 × pixel size = 14.06 mm

128 × 128 matrix


FWHM 3 × 300 mm / 128 = 7.03 mm

The values calculated in Example 20-1 represent the approximate levels of imaging system resolution that could be supported without loss of imaging resolution for the specified image and matrix sizes. The practical effects of undersampling depend as well on the information contained in the image and whether it has a significant amount of actual spatial frequency content near the resolution limit of the imaging device. Practical experimentation sometimes is required to determine this for a particular type of imaging procedure.