Patients with implanted pacemakers or automatic implantable cardioverter defibrillators (AICDs) are commonly seen in the emergency department (ED). Fortunately, the increased reliability of these devices has precluded a marked increase in patients with true emergencies related to device malfunction, but such patients clearly have serious underlying medical problems that must be considered. Pacemaker complications are not uncommon, with rates ranging from 2.7% to 5%. Many pacemakers fail within the first year. AICD complication rates, including inadvertent shocks, occur in up to 34% of patients with the device. The basic evaluation and treatment of patients with cardiac complaints are not substantially different in patients with pacemakers and AICDs than in those without. However, a general knowledge of the range of problems, complications, and techniques for evaluating or inactivating pacemakers or AICDs is important for emergency clinicians. These devices are complicated and emergency physicians are not expected to be experts in all complications; therefore appropriate consultation is often necessary depending on the clinical situation.

Pacemaker Characteristics

In essence a pacemaker consists of an electrical pulse–generating device and a lead system that senses intrinsic cardiac signals and then delivers a pulse. The pulse generator is hermetically sealed with a lithium-based battery device that weighs approximately 30 g and has an anticipated lifetime of 7 to 12 years. A semiconductor chip serves as the device's central processing unit. The generator is connected to sensing and pacing electrodes that are inserted into various locations in the heart, depending on the configuration of the pacemaker. Newer models are programmable for rate, output, sensitivity, refractory period, and modes of response. They can be reprogrammed radiotelemetrically after implantation.

Pacemakers are classified according to a standard five-letter code developed by the North American Society of Pacing and Electrophysiology/British Pacing and Electrophysiology Group ( Table 13.1 ). Known as the NBG code, it consists of five positions or digits. The first letter designates the chamber that receives the pacing current; the second, the sensing chamber; and the third, the pacemaker's response to sensing. The fourth letter refers to the pacemaker's rate modulation and programmability, and the fifth describes the pacemaker's ability to provide an antitachycardia function. Whereas standard pacemakers generally do not have an antitachycardia function, AICDs do have this capability and overdrive pacing is the device's first response to tachycardia. In normal practice, only the first three letters are used to describe the pacemaker (e.g., VVI or DDD).

TABLE 13.1
North American Society of Pacing and Electrophysiology/British Pacing and Electrophysiology Group Generic Pacemaker Code (NBG Code)
I II III IV V
Chamber Paced Chamber Sensed Response to Sensing Rate Modulation and Programmability Antitachycardia Features
0—None 0—None 0—None 0—None 0—None
A—Atrium A—Atrium I—Inhibited I—Inhibited P—Antitachycardiac pacing
V—Ventricle V—Ventricle T—Triggered M—Multiple S—Shock
D—Dual D—Dual D—Dual C—Communicating D—Dual
R—Rate modulation

Pacemaker wires are embedded in plastic catheters. The terminal electrodes, which may be unipolar or bipolar, travel from the generator unit to the heart via the venous system. In a unipolar system, the lead electrode functions as the negatively charged cathode, and the pulse generator case acts as the positively charged anode into which electrons flow to complete the circuit. The pulse generator casing must remain in contact with tissue and be uninsulated for pacing to occur. In the case of bipolar systems, both electrodes are located within the heart. The cathode is at the tip of the lead, and the anode is a ring electrode roughly 2 cm proximal to the tip. Bipolar leads are thicker, draw more current than unipolar leads, and are commonly preferred because of several advantages, including a decreased likelihood of pacer inhibition as a result of extraneous signals, and decreased susceptibility to interference by electromagnetic fields.

The typical entry point for inserting the leads is the central venous system, which is generally accessed via the subclavian or cephalic vein. The terminal electrodes are placed either in the right ventricle or in both the right ventricle and the atrium under fluoroscopic guidance. Proper lead placement, sensing, and pacing thresholds are assessed with electrocardiograms (ECGs). The typical radiographic appearances of implanted pacemakers and implantable cardioverter-defibrillators to include an example of lead fracture are depicted in Fig. 13.1A–E .

Figure 13.1, A, Various radiographs of an implanted pacemaker and automatic implantable cardioverter-defibrillator showing battery and lead wires. Posteroanterior (PA; A1 ) and lateral ( A2 ) chest radiographs demonstrate a biventricular pacing system. There are three leads—the first is positioned in the right atrium, the second is in the right ventricular apex, and the third courses posteriorly in the coronary sinus and into the posterolateral cardiac vein. B, PA chest radiographs of a dual-chamber pacemaker. B1, The ventricular lead is passing through an atrial septal defect into the left ventricle. B2, The lead is repositioned in the right ventricular apex. C, A dual-chamber implantable cardioverter-defibrillator with active fixation leads has been implanted via a transvenous approach to place the atrial lead in the systemic venous atrium and the ventricular lead across the baffle into the morphologic left ventricle. D, PA chest radiograph of a patient with a dual-chamber pacemaker ( D1 ). The atrial lead, originally positioned in the right atrial appendage, is clearly no longer positioned in the right atrial appendage. A lateral view ( D2 ) also shows definite dislodgement of the atrial lead. E, Close-up view of a portion of the PA chest radiograph of a patient with a single-chamber pacemaker. The lead has fractured (a subtle finding) at the point where it passes below the clavicle (arrow). The device showed intermittent ventricular failure to capture and intermittent failure to output on the ventricular lead. Impedance was intermittently measured at more than 9999 Ω. Inset, Diagram of the fracture site.

The pacemaker is typically programmed to pace at a rate of 60 to 80 beats/min. A significantly different rate usually indicates malfunction. When the battery is low, the rate generally begins to drop and gets slower as the battery fades. Sensing of intracardiac electrical activity is a combination of recognizing the characteristic waveforms of P waves or QRS complexes while discriminating them from T waves or external interfering signals, such as muscle activity or movement. The pacing electrical stimulus is a triphasic wave consisting of an intrinsic deflection, far-field potential, and an injury current, which typically delivers a current of 0.1 to 20.0 mA for 2 msec at 15 V.

Pacemakers have a reed switch that may be closed by placing a magnet over the generator externally on the chest wall; this inactivates the sensing mechanism of the pacemaker, which then reverts to an asynchronous rate termed the magnet rate . Essentially, the magnet turns the demand pacemaker into a fixed-rate pacemaker. The magnet rate is usually, but not always the same as the programmed rate.

Several innovations in rate regulation have been incorporated into some pacemakers. When present, the hysteresis feature causes pacing to be triggered at a rate greater than the intrinsic heart rate. When the hysteresis feature is used in a single-chamber ventricular pacemaker, it is designed to maintain atrioventricular (AV) synchrony at rates that are lower than what would be normal for a ventricular-paced rhythm alone. To illustrate, were the hysteresis feature of the pacemaker set at 50 beats/min, an intrinsic rate lower than 50 beats/min would trigger ventricular pacing. Unlike a standard ventricular pacemaker, the hysteresis feature might be set to offer a ventricular pacing rate at 70 beats/min or greater once the pacer is triggered.

Rate modulation by sensor-mediated methods is an additional feature triggered and mediated by a sensed response to various physiologic stimuli. The primary application for this rate modulation feature is in patients with pacemakers who continue to engage in vigorous physical activity. When present, the rate regulation feature is engaged and modulated through motion sensors installed within a pulse generator device, with a corresponding increase or decrease in the pacing rate depending on the degree of motion sensed by the pacemaker device. Other physiologic sensors that may be installed as part of the pacemaker system include those designed to sense minute ventilation, the QT interval, temperature, venous oxygen saturation, and right ventricular contractions. The latter sensors generally require that additional leads be placed.

One of the newest innovations in cardiac pacing is cardiac resynchronization therapy (CRT), also known as biventricular pacing. In addition to a conventional right ventricular endocardial lead, CRT also employs a coronary sinus lead for left ventricular pacing. A right atrial lead may also be included. The primary objective of CRT is to restore left ventricular synchrony, primarily in patients with dilated cardiomyopathy and widened QRS complex. This condition results primarily from the presence of a left bundle branch block. In such situations CRT may improve left ventricular function, and thus cardiac output.

Characteristics of AICDs

The basic components of an AICD, including sensing electrodes, defibrillation electrodes, and a pulse generator ( Fig. 13.2 ), can be seen on a chest radiograph. Transvenous electrodes have obviated the previous need for surgical placement. They are inserted into the pectoralis muscle. Many transvenous systems consist of a single lead containing a distal sensing electrode and one or more defibrillation electrodes in the right atrium and ventricle. Leads are inserted through the subclavian, axillary, or cephalic vein into the right ventricular apex. The left side is preferred because of a smoother venous route to the heart and a more favorable shocking vector. In an effort to improve the efficiency of defibrillation, an additional defibrillation coil may be used. Various placements of AICDs are demonstrated in Fig. 13.3 .

Figure 13.2, A, Automatic implantable cardioverter-defibrillator. B, Implantable pacemaker.

Figure 13.3, Diagrammatic demonstration of various automatic implantable cardioverter defibrillator (AICD) configurations. A, Single-chamber AICD. B, Dual-chamber AICD. C, Biventricular AICD.

The pulse generator is a sealed titanium casing that encloses a lithium–silver–vanadium oxide battery. It has voltage converters and resistors, capacitors to store charge, microprocessors and integrated circuits to control analysis of the rhythm and delivery of therapy, memory chips to store electrographic data, and a telemetry module. Although a pacemaker can draw the voltage required for function from its component battery, the energy needed for defibrillation requires a battery that is prohibitively large. To circumvent this problem, an AICD contains a capacitor that maximizes the voltage required by transferring energy from the battery before discharge. To achieve the energy required, AICDs use capacitors that are charged over a period of 3 to 10 seconds by the battery and then release this energy rapidly for defibrillation. The maximal output is 30 J in most units and 45 J in higher-energy units. This energy is high enough that a discharge is very obvious and often distressing to the patient.

Most AICDs use a system in which the pulse generator is part of the shocking circuit, often described as a “can” technology, and most of them have a dual-coil lead with a proximal coil in the superior vena cava and a distal coil in the right ventricle. Current flows in a three-dimensional configuration from the distal coil to both the proximal coil and the generator. This dispersion of the electrical field increases the likelihood of depolarizing the entire myocardium at once, thereby leading to successful defibrillation.

AICDs may have the same programming capabilities as pacemakers and can be single chambered, dual chambered, or used with the aforementioned CRT. Single-chamber devices have only a right ventricular lead. They often have difficulty identifying atrial arrhythmias, which can result in inappropriate defibrillation of atrial tachycardias. Dual-chamber AICDs have right atrial and right ventricular leads and improved ability to discriminate rhythms. In most studies, dual devices have been found to offer improved discrimination between ventricular and supraventricular arrhythmias, thus decreasing inappropriate shocks as a result of rapid supraventricular rhythms or physiologic sinus tachycardia. Of AICDs implanted annually 27% are single chamber, 32% dual chamber, and 41% are CRT systems. In patients requiring both AICD and pacemaker functions, both these devices are placed together. The advent of technology has allowed placement of a single device that can perform both pacemaker and defibrillator functions.

AICDs use a combination of antitachycardia pacing, low-energy cardioversion, defibrillation, and bradycardiac pacing in a combination also known as tiered therapy . They are programmed with specific algorithms that identify and treat specific rhythms. Ventricular arrhythmias may initially be converted (or undergo attempts at conversion) with antitachycardiac pacing as opposed to immediate defibrillation. This overdrive pacing may terminate the rhythm without the need for electrical defibrillation in up to 90% of events. It is most successful for terminating monomorphic ventricular tachycardia with a rate of less than 200 beats/min. Overdrive pacing is better tolerated by patients than cardioversion and reduces the risk for inducing atrial fibrillation. These events may be silent, not felt by the patient, and discovered only by interrogating the device.

If unsuccessful, the next intervention may be low-energy cardioversion (<5 J). The device may be programmed to very low levels of electricity that are better tolerated by the patient. This works best for ventricular rates higher than 150 and lower than 240 beats/min. This may be followed by a high-energy defibrillation. Traditionally, the energy level of the first shock is set at least 10 J above the threshold of the last defibrillation measured. If the first shock fails a backup shock may be required, but this may induce or aggravate ventricular arrhythmias (see the later section on Pacemaker-Mediated Tachycardia ). Unlike the proarrhythmic effects of medication, these arrhythmias are almost never fatal, although they may be associated with increased morbidity. Currently used biphasic waveforms have improved defibrillation thresholds.

This tiered approach obviates the need for unnecessary energy requirements. The devices also have antibradycardiac pacing that allows these patients to have one device instead of separate units. Additional complications associated with AICDs that have antibradycardiac pacing algorithms include a tendency toward oversensing, increased current drain, potential detection problems, and an increased incidence of hardware and software design problems. At the time of insertion, the amount of energy required for various AICD functions, such as the defibrillation threshold, is determined for any given patient, and output and sensing functions can be adjusted by reprogramming as needed.

In 2012 the Food and Drug Administration approved the use of a subcutaneous defibrillator. This consists of a subcutaneous lead with a shock coil flanked by two sensing electrodes; this coil is tunneled along the parasternal margin from the pulse generator that is implanted in a subcutaneous pocket. It is placed by anatomic landmarks and as such no fluoroscopy is required. It delivers an 80-J biphasic shock across the precordium. Fig. 13.4 is a radiograph of a patient with a subcutaneous defibrillator.

Figure 13.4, Chest radiograph demonstrating a subcutaneous defibrillator.

The subcutaneous implantable device may be an alternative to patients who have difficult vascular access or who are at risk for transvenous complications. These could include the pediatric populations, patients with congenital disease, complicated vascular anatomy, patients at high risk for infection, or patients on dialysis. The notable limitations of the subcutaneous internal defibrillator when compared with conventional devices is the inability to provide long-term bradycardia pacing or antitachycardiac pacing. Patients that need resynchronization therapy are also unsuitable candidates.

Left Ventricular Assist Devices

As the management of heart failure continues to evolve, emergency physicians may encounter patients who have surgically implanted ventricular assist devices (VADs). VADs are implanted pumps used to assist a failing ventricle. Although they have been used in the right ventricle and for biventricular support, they are primarily used as a left ventricular assist device or LVAD. The first-generation of these devices utilized a pulsatile flow pump. The second and third devices provide continuous flow through the use of axial flow or centrifugal pumps. These devices are smaller and more durable, with higher energy efficiency than the first generation models, which are no longer in use.

Initially used as a bridge to heart transplantation, they are now also being used as a bridge to recovery of heart function (such as in patients with myocarditis) and for permanent (destination) therapy. The LVAD consists of an inflow cannula that is placed in the apical part of the ventricle and allows blood to flow to a pump. The pump is attached to the outflow cannula, which is placed in the ascending aorta. The device is connected through a percutaneous lead or driveline to batteries located outside the body. Figs. 13.5 and 13.6 demonstrate the components of a left ventricular device.

Figure 13.5, HeartWare left ventricular device basic components and function.

Figure 13.6, The HeartMate II continuous flow device receives blood from the left ventricular apex and returns to the ascending aorta. The pump is connected with a percutaneous lead to an eternal controller and batteries worn by the patient.

Indications for Placement of Implantable Pacemakers and Aicds

The most common indication for placement of a cardiac pacemaker is for the treatment of symptomatic bradyarrhythmias. Roughly 50% of pacemakers are placed in such patients for the treatment of sinus node dysfunction (sick sinus syndrome). Other diagnoses include symptomatic sinus bradycardia, atrial fibrillation with a slow ventricular response, high-grade AV block (including Mobitz type II and third-degree AV block), tachycardia-bradycardia syndrome, chronotropic incompetence, and selected prolonged QT syndromes. Though not classified as absolute indications, pacemakers are sometimes placed for the treatment of severe refractory neurocardiogenic syncope, paroxysmal atrial fibrillation, and hypertrophic or dilated cardiomyopathy.

In recent years, CRT has emerged as a primary approach for patients with severe diastolic dysfunction and a low left ventricular ejection fraction (LVEF). Commonly such patients manifest low-grade AV blocks and left bundle branch block. The resultant delay in left ventricular conduction often results in corresponding biomechanical delays in ventricular contraction, which in turn cause a further decrement in cardiac output and worsening congestive heart failure. Such prolongation may occur in as many as 33% of patients with advanced heart failure. This electromechanical “dyssynchrony” has been associated with increased risk for sudden cardiac death.

CRT comprises atrial-synchronized, biventricular pacemaking, which overcomes the atrial and ventricular blocks while optimizing both preload and LVEF. Clinical trials and systematic reviews have confirmed the efficacy of CRT, with decrements in mortality of 22% to 30%, in addition to improved LVEF and quality of life. It is therefore likely that emergency physicians will see the CRT configuration with increasing frequency in patients with implanted pacemakers and AICDs.

AICD technology is used principally for both primary and secondary prevention in patients at risk for sudden death. Primary prevention is an attempt to avoid a potentially malignant ventricular arrhythmia in patients identified as being at high risk. Secondary prevention is for patients who have already had a ventricular arrhythmia and are at risk for further events. In addition, AICDs are implanted for a number of other congenital or familial cardiac conditions.

Pacemaker and AICD Response to Magnet Placement

In the clinical setting, placement of a magnet over the pulse generator of a pacemaker is a technique that can be used either diagnostically or therapeutically by the emergency clinician. Application of a magnet is used to suspend antitachyarrhythmia detection and therapy of an AICD without changing pacing mode, or to produce asynchronous (fixed rate) pacing of a pacemaker. Hence, a magnet may change a demand pacemaker (synchronous) to a fixed rate (asynchronous) pacemaker. However, it is important to note that each pacemaker is programmed to respond in a specific fashion to a magnet as determined by the manufacturer. The response to magnet placement may not only vary by manufacturer, but also by model and by the particular mode in which the pacemaker is currently operating. Some devices (Boston Scientific, St. Jude, Biotronic devices) may be programmed to disable the magnet function and hence not respond to magnet application. Some leadless pacemakers do not initiate asynchronous pacing in response to a magnet application. In most cases manufacturers set the asynchronous (fixed rate) baseline pacing rate in a range approximating 70 beats/min. An indicator of aging of the pacemaker and weakening of the battery is that this asynchronous baseline pacing rate will decrease over time as the battery approaches the point at which replacement is required.

Keeping these provisions in mind, there are standard responses that the provider might expect to see in most circumstances. In the case of single-chamber ventricular pacemakers, the response will most likely be asynchronous pacing (V00). In the case of dual-chamber pacemakers, placement of a magnet usually results in dual-chamber asynchronous pacing (D00). In either case, it is important for the clinician to note that placement of a magnet over the pacemaker pulse generator will not turn the pacemaker off.

Placing a magnet over any of the currently available AICD models will temporarily disable tachyarrhythmia intervention ( Fig. 13.7 ). An ECG should be obtained before and after magnet placement for comparison ( Fig. 13.8 ). Most commercially available pacer magnets are 7 cm in size and can be used with most implantable devices. Each of the present models may have a slightly different response to the magnet. The magnetic field closes a reed switch in the generator circuit that will disable recognition of tachyarrhythmias and subsequent firing of the device. There may be a variety of tones (continuous, intermittent, or silent) during activation or inactivation with the magnet, which are dependent on the manufacturer. Some devices may be programmed to not respond at all. After the desired effect is obtained, the magnet should be secured to maintain inactivation.

Figure 13.7, A, This patient suffered from 13 discharges of his automatic implantable cardioverter-defibrillator over the course of 1 hour. In the emergency department, he was noted to have a discharge while in sinus rhythm, and a magnet (arrow) was placed over the device to deactivate it. B, An electrophysiology specialist came to the emergency department to interrogate the device. Note that newer devices can communicate wirelessly. C and D, Interrogation and intracardiac electrocardiograms were consistent with fracture of the right ventricular lead. The device was deactivated and replaced shortly thereafter.

Figure 13.8, A, Electrocardiogram (ECG) of a patient with a nonfiring pacemaker. The intrinsic cardiac rate is 80 beats/min, and no pacemaker activity is seen. B, ECG of the same patient with a magnet applied over the pacemaker, which produced a paced rhythm. Pacer spikes are evident (arrows), and the magnet rate is 85 beats/min. Note the left bundle branch bundle typical of a pacer lead in the right ventricle.

Pacemaker and AICD patients should carry an identification card that includes information regarding manufacturer, model type, and lead system, in addition to a 24-hour emergency number to allow rapid identification of the model when it is necessary to inactivate the device. In lieu of the availability of a device identification card, the general type, polarity, and number of ventricles involved with the implanted device may be inferred accurately by viewing an overpenetrated anteroposterior chest radiograph.

If a patient with an AICD has a ventricular arrhythmia, the assumption should be made that the device is inoperable and standard advanced cardiac life support (ACLS) protocols should be used to stabilize the patient.

Of further note, in some obese patients or those with heavily developed chest wall musculature, the magnetic field emitted by a single magnet device may not be strong enough to elicit the desired effect on the implanted device. In such cases the clinician may find greater efficacy by using two magnets, one on top of the other.

Clinical Evaluation of Patients With Implanted Pacemakers and Aicds

History

Although patients who go to the ED because of implantable pacemaker–related issues may have one or more of several complaints, those with AICD-related issues are generally seen because their device has discharged. They will often describe a sensation of being kicked or punched in the chest, and the sensation is not subtle. In fact, some patients live in fear of the shock after having experienced it previously, and this is one reason for removal of the device. Ask the patient about the number of discharges and associated symptoms, including chest pain, shortness of breath, lightheadedness, palpitations, syncope, extremity edema (raising concern for congestive heart failure or lower extremity deep vein thrombosis), or dyspnea on exertion. In addition, elicit general symptoms such as fever, chills, nausea, or vomiting, which could be indicative of infection. Inquire about medication history. Ask about the specific implanted device that they possess. Most pacemaker and AICD patients should have an identification card on their person that will identify the manufacturer, model number, lead system, and a 24-hour emergency contact number. Sophisticated information and the prior electrical events and settings of the device can be ascertained in the ED by simply placing an external interrogating device over the unit.

When a patient with an LVAD arrives in the ED, the provider should identify who the patient's cardiologist and cardiothoracic surgeon are. In most instances contact should be made with these providers and the patient may need to be transferred to a center with expertise in these devices.

Physical Examination

First, assess for airway patency, adequate ventilation, and cardiovascular status. The patient's mental status may also be an important clue to the severity of the symptoms. Perform an appropriate physical examination with emphasis on examining the heart and lung to seek murmurs, pericardial friction rubs, and evidence of pulmonary effusion or other abnormalities. Inspect the pacemaker or AICD site for erythema or edema. Palpate the skin for evidence of obvious lead abnormalities. Examine the extremities for evidence of edema or erythema.

There are several unique characteristics to consider in patients with ventricular assist devices. On physical exam, the provider should listen to the chest to determine if an audible “hum” is present. Patients lack a palpable pulse given the device's continuous flow and because their cardiac function at baseline is poor. The blood pressure should be obtained by using a Doppler on either the radial or brachial artery. Inflate the cuff until you can't hear flow. The cuff pressure should then be released, listening for the presence of flow. There is some degree of controversy about whether this is mean arterial pressure (MAP) or systolic pressure, however. When in doubt, an arterial line would be the most accurate method to identify MAP. MAP is usually kept between 70 to 80. Pulse oximetry is also unreliable and a true measurement should be obtained with an arterial blood gas or arterial line.

Radiography

The imaging modality of choice, at least initially, in a patient with complaints related to an implanted pacemaker or AICD is the plain chest radiograph. If the patient's stability is in question, obtain a portable anteroposterior film. In addition to the standard cardiac, pulmonary, vascular, and skeletal evaluation, this study will probably confirm the location of the pulse generator case, as well as the current location of any leads. Survey the radiograph for evidence of lead fracture or displacement (see Fig. 13.1 E ). Compare with previous chest films. If the patient does not have a device identification card in immediate possession, take an overpenetrated radiograph to look for a radiopaque marker identifying the model type. Fig. 13.9 demonstrates a radiograph of a patient with an LVAD.

Figure 13.9, Heart-Mate II model of left ventricular assist device (LVAD) appearance in A, posterior–anterior and B, lateral chest X-ray, C, computerized tomography, and D, ultrasound examination. ICD, Implantable cardioverter defibrillator; LV, left ventricle; LA, left atrium.

Electrocardiography

Perform an ECG to seek evidence of pacing, ectopy, and ST-segment or T-wave abnormalities. A comparison ECG may be helpful. If an AICD shock has been delivered, the shock itself can cause transient electrocardiographic changes, and waiting for several minutes to repeat the ECG may identify whether the changes are caused by the discharge or an ongoing disease process.

Cardiopulmonary Resuscitation, ACLS Interventions, and External Cardiac Defibrillation in Patients With Implanted Pacemakers or AICDs or LVADS

In general, ACLS interventions may be performed safely and effectively in patients with pacemakers and AICDs when indicated. Cardiopulmonary resuscitation (CPR) can usually be performed in standard fashion. If an AICD is present, rescuers may notice mild electrical shocks while performing CPR; these shocks are harmless to the rescuer. If the AICD shocks are impeding rescuer performance of CPR, or if supraventricular tachycardias are noted during resuscitation, disable the AICD by applying a magnet over the corner of the device from which the leads emerge. This location is generally found easily by palpation but may be located blindly by slowly relocating the magnet until AICD activity ceases.

External cardiac defibrillation may be performed safely in patients with pacemakers and AICDs with the standard expected efficacy; however, it is recommended that external paddles or defibrillator pads be placed at a location approximately 10 cm distant from the pulse generator if possible. A transcutaneous cardiac pacemaker may also be used in similar fashion with a recommendation that the pacing pads be placed in anatomically appropriate locations but preferably at a distance of 10 cm from the pulse generator.

Placing the external defibrillation or transcutaneous pacemaker pads in an anteroposterior configuration is advised because this configuration may circumvent energy shunting and shielding. Every attempt should be made to avoid application of the defibrillators directly over the device. Use of the lowest possible energy setting for cardioversion or defibrillation is recommended. If available, biphasic cardioverter defibrillators are further suggested. In the event of successful resuscitation and return of spontaneous circulation, the pacemaker or AICD should be interrogated expeditiously by a cardiologist or electrophysiologist to ensure that no damage was sustained as a result of the resuscitation effort.

Regarding pharmacologic adjuncts, amiodarone has been reported to be more effective for the treatment of potentially lethal arrhythmias in the setting of implanted devices. Antiarrhythmic medications may be required for a resistant malignant rhythm when the AICD is functioning properly but the arrhythmia persists ( Fig. 13.10 G ).

Figure 13.10, A, Patient with an automatic implantable cardioverter defibrillator (AICD) who experienced multiple discharges from an AICD because of recurrent ventricular tachycardia (VT). B, Electrocardiogram (ECG) of VT in the presence of an underlying tachycardia as the possible cause. C, A rhythm strip demonstrates persistent polymorphic (torsades de pointes) VT. D, After attempts at overdrive pacing failed, the AICD appropriately discharges with temporary termination of VT. E, Interrogation of the device uncovers a history of more than 20 episodes of recurrent VT within 30 minutes. This patient was terrified of a subsequent shock because he could sense the VT and knew of the impending shock. In addition to antiarrhythmic medication, a quiet dark room and aggressive sedation are suggested to reduce catecholamine levels. F, ECG after medical therapy with multiple medications, including magnesium. Metoprolol seemed to be the deciding factor in terminating the VT in this case. G, Current suggested medical therapy for VT in the presence of an appropriately functioning AICD.

Several additional considerations are unique to the setting of ACLS in patients with a pacemaker or AICD. In cases of acute myocardial infarction involving areas of the myocardium in contact with the pacemaker leads, the implanted pacemaker may experience operative failure. Therefore, maintain a high level of suspicion for the potential requirement for supplemental transcutaneous or transvenous cardiac pacing.

LVAD patients that go into cardiac arrest offer some unique challenges. Ensure that the machine is connected, the battery charged, and the machine has an audible hum. The lack of peripheral pulses also make the use of ACLS problematic. In the case of arrest, follow the standard Advanced Cardiac Life Support algorithms. There is concern that standard CPR may dislodge the cannula or damage the unit's outflow conduit, but there have also been reports that this is not as harmful as previously thought. There was also a report of an emergency physician who replaced the driveline connections in a patient who had cut the wires resulting in subsequent resumption of the pump function.

In the postresuscitation setting involving a patient with an implanted pacemaker or AICD, it is important to maintain a higher index of suspicion for device lead fracture or disruption resulting from CPR. Finally, in the postresuscitation phase the clinician should closely watch for the development of pneumothorax, hemothorax, pericardial effusion, or other aforementioned pathophysiologic processes that could adversely affect the function of the implanted device.

Complications and Malfunction of Implanted Pacemakers

Complications associated with pacemakers are listed in Box 13.1 . In addition, patients with previously implanted and otherwise stable pacemakers may experience complications related to direct or indirect trauma affecting the pulse generator or leads. Major complications of pacemaker placement or those caused by subsequent injuries that the emergency clinician might encounter include: local or systemic infections resulting from pacemaker placement, thrombophlebitis involving the transvenous route of the pacemaker leads, a venous thromboembolic event, pneumothorax or hemothorax, pericarditis, air embolism, localized hematoma interfering with pacemaker operation or sensing, lead dislodgement, cardiac perforation, hemopericardium with possible progression to cardiac tamponade, and development of the phenomenon known as pacemaker syndrome . This condition is often seen in patients with single-chamber ventricular pacemakers who have an underlying component of congestive heart failure. It is believed to be a consequence of the loss of AV synchrony resulting from the ventricular pacing, and may be manifested as vertigo, syncope, hypotension, and signs specific to the exacerbation of congestive heart failure.

Box 13.1
Complications of Permanent Pacemakers

Failure to Pace (No Pacemaker Activity Present)

  • Lead fracture

  • Lead disconnection

  • Battery depletion

  • Component failure

  • Oversensing

  • External interference

You're Reading a Preview

Become a Clinical Tree membership for Full access and enjoy Unlimited articles

Become membership

If you are a member. Log in here