Positron Emission Tomography and Computed Tomography Technique and Instrumentation


Technical Aspects

Positron emission tomography and computed tomography (PET/CT) represents the successful technical combination of multidetector computed tomography (MDCT) and PET into a single scanner. PET with the fluorine-18 (18F)–labeled glucose analog fluorodeoxyglucose (FDG) provides metabolic imaging of tissues, both normal and diseased. FDG-PET provides valuable qualitative and quantitative metabolic information for both diagnosis and management. PET has been shown to be of value in diagnosing and staging malignant tumors as well as in follow-up oncologic imaging after surgical, radiation, or chemotherapeutic treatment. However, the sensitivity of PET in detecting hypermetabolic foci is compromised in large part by the low background FDG uptake when attempting accurate anatomic localization.

CT provides rapidly acquired data sets of high spatial resolution giving multiplanar information regarding the morphologic features and attenuation values of both normal anatomy and pathologic lesions. A major weakness of CT, however, is its dependence on changes in the size, shape, or attenuation values of a structure to detect pathologic processes.

Separate CT and PET scans of the same patient acquired on different scanners certainly can be aligned using a number of available software methods, but the algorithms are often labor intensive and, particularly in the abdomen, may fail to provide a satisfactory overlap. However, combined imaging with PET/CT allows for the precise structural information provided by CT to accurately locate the hypermetabolic foci identified with PET and thereby improve the overall diagnostic performance ( Figure 12-1 ). The information from combined PET/CT also facilitates identification of physiologic from pathologic uptake and diminishes the false-negative and false-positive interpretations of the individual components. Indeed, it has been shown that PET/CT is more accurate for many oncologic entities than either CT or PET alone. Furthermore, PET has the potential to be used with radiotracers different from FDG to provide other useful biological information, whereas CT provides useful information about the behavior of orally and intravenously administered contrast agents.

Figure 12-1, Value of precise localization and complementary strengths. Appendiceal adenocarcinoma. Markedly dilated appendix with thick wall and communication with cecum shows increased fluorodeoxyglucose uptake (arrows) along its wall. Without the positron emission tomography (PET) findings, the computed tomography (CT) finding might have been overlooked with the dilated appendix mimicking a large bowel loop. Without the CT findings, the PET activity could not have been precisely localized and possibly may have been misinterpreted as physiologic bowel activity.

Instrumentation Considerations

The first PET/CT prototype scanner was revealed in 1998. Subsequently, the first clinical scanners appeared in 2001, and by early 2007 more than 800 combined PET/CT scanners had been installed in clinical institutions worldwide. Essentially all new PET machine sales are now as PET/CT scanners.

The automatic and more accurate coregistration of the structural and metabolic data is just one of the inherent advantages of hybrid PET/CT machines. PET-only whole-body scans traditionally required up to 45 to 60 minutes to complete. The transmission scan required for attenuation correction to improve the qualitative and quantitative accuracy of PET is a major contributor to this long scan time. However, CT-based attenuation correction is significantly quicker than traditional PET transmission methods and can provide almost noise-free information and allows for total PET/CT scanning durations of 20 to 30 minutes or less. This greatly reduced scanning time enhances patient comfort and convenience and also allows a greater patient throughput. The current emission PET scan time may now also be decreased by the use of time-of-flight (TOF) technology, which shortens scan time and improves signal-to-noise ratio.

Feasible TOF now can be applied owing to the newer fast scintillators with high stopping power, such as lutetium oxy­orthosilicate (LSO) and LYSO (which is LSO with the addition of a small percentage of yttrium). TOF makes use of the ability of the new scanners to measure the arrival time difference between the two positron annihilation-generated photons reaching the detectors to within a certain resolution. Recent detector and scintillator developments allow sub-nanosecond coincidence timing resolution, which provides a rapid, TOF-based and back-projection–free, three-dimensional (3D) reconstruction algorithm that, combined with real-time data acquisition and a quick detector encoding scheme, permits high-quality images to be obtained in a significantly reduced time, particularly in nonobese patients. The stability of these new scanners in practice has yet to be confirmed; and although encouraging and interesting, the precise clinical contribution of TOF PET has yet to be determined. Certainly, high spatial resolution PET detectors appear to optimize the conspicuity of FDG uptake in small lesions.

From the late 1980s until recently, the bismuth germanate (BGO) block detector was the standard for PET. However, the introduction of new faster scintillators with greater light output and shorter decay time than BGO, such as gadolinium oxyorthosilicate (GSO) and LSO (and LYSO), has recently improved the performance of PET scanners for clinical imaging. The faster scintillators decrease the coincidence timing window and thus the random coincidence rate. The improved light output of the new scintillators makes the energy resolution more accurate by reducing the statistical uncertainty of the measurement. This higher light output also improves the positioning accuracy of a block detector, with potential spatial resolution benefits.

The block design developed by Casey and Nutt has been the basic detector component in all multi-ring PET scanners for the past 2 decades. The first multi-ring PET scanners incorporated septa between the detector rings to shield the detectors from scattered photons out of the transverse plane, thus not allowing 3D reconstruction algorithms. The first multi-ring PET scanners with retractable septa that included the capability to acquire data in either the 2D or 3D mode appeared in the early 1990s. The advantages of 3D imaging for the brain are widely accepted, but 3D imaging for the rest of the body has posed more problems. Recently, however, statistically based reconstruction algorithms, more accurate scatter correction methods, and faster scintillators have significantly improved the achievable whole-body 3D imaging quality. However, 2D imaging still may be more suitable for large patients. Increased axial coverage makes better use of the emitted radiation for a given volume of scintillator. For most PET/CT scanners, axial PET coverage is approximately 15 cm, ranging up to a recently extended 21.6 cm, with resulting imaging advantages.