Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
The conduct of anaesthesia is planned after obtaining details of the surgical procedure and medical condition of the patient. All patients should be visited preoperatively for anaesthetic assessment, including review of the results of relevant investigations. This is also an important opportunity to form a rapport, gain patients’ trust and answer any questions about their anaesthetic and postoperative pain relief. Preoperative assessment and selection of appropriate premedication are discussed in Chapter 19 .
Before commencing anaesthesia and surgery, the anaesthetist must be satisfied that there is functioning equipment, the correct monitoring is attached and set up, the trained assistant is present and the WHO patient safety checklist has been completed (see Chapter 18 ).
Anaesthetic equipment should be up-to-date and maintained regularly and the instruction manuals available and accessible. An equipment check must be performed before an operating theatre session begins, because a common cause of problems is the use of a machine that has not been checked properly and which malfunctions. An adequate check of anaesthetic apparatus is an integral part of good practice; failure to check the anaesthetic equipment properly may amount to malpractice.
Operating theatre staff usually carry out checks when setting up an operating theatre for use or after apparatus has been serviced or repaired, but the ultimate responsibility for ensuring that the apparatus is safe for its intended use rests with the anaesthetist. Sophisticated tests may have been performed after major servicing, but key control settings may have been altered, and it is essential that the anaesthetist checks that the equipment is in proper working order and ready for clinical use. The final check is the sole responsibility of the anaesthetist who is to use the machine; it cannot be delegated to any other person. There is no justification for proceeding with an anaesthetic when faults have been identified in the equipment. If there is no record of an adequate preoperative check of equipment and a problem occurs as a result of equipment failure, it is very difficult to defend an allegation of negligence.
At its most basic, the function of an anaesthetic machine is to enable the anaesthetist to administer oxygen under pressure without leaks. If all else fails, this allows the anaesthetist to maintain cerebral oxygenation. Anaesthetic apparatus should be checked before the start of each operating session in a logical sequence, as recommended on the Association of Anaesthetists checklist ( Table 22.1 ). Further checks should be undertaken between cases ( Box 22.1 ). The primary intention of the anaesthetic machine check is to ensure that it is safe to use and will deliver gases under pressure without leaks. These checklists are available as laminated cards intended to be attached to all anaesthetic machines in the UK and Ireland.
Checks at the start of every operating session Do not use this equipment unless you have been trained |
|
Check self-inflating bag available | |
Perform manufacturer's (automatic) machine check | |
Power supply |
|
|
|
|
|
Gas supplies and suction |
|
|
|
|
|
|
|
|
|
|
|
Breathing system |
|
|
|
|
|
|
|
|
|
Ventilator |
|
Scavenging |
|
Monitors |
|
|
|
Airway equipment |
|
RECORD THIS CHECK IN THE PATIENT RECORD | |
Don't forget! |
|
|
|
|
|
|
|
|
Whole system patent and leak-free using two-bag test (see Box 22.2 )
Vaporisers – fitted correctly, filled, leak-free, plugged in (if necessary)
Alternative systems (e.g. Bain, T-piece) – checked
Correct gas outlet selected
Working and configured correctly
Full range required working, with spares
Clean and working
In most industries in which complex equipment is used, full training is provided for users. It is not acceptable for anaesthetists to assume that they intuitively understand an anaesthetic machine they have not used before. Those new to the specialty require detailed instruction and training in the use of anaesthetic equipment, but even experienced anaesthetists need tuition in the use of new equipment.
The early use of an alternative means of ventilation, namely a self-inflating bag that does not rely on a source of oxygen to function, may be lifesaving. A self-inflating bag must be immediately available in any location where anaesthesia might be given. An alternative source of oxygen should also be readily available.
Modern anaesthesia workstations may perform many of the following checks automatically during start-up. Anaesthetists must know which are included and ensure that the automated check has been performed.
The anaesthetic workstation and relevant ancillary equipment must be connected to the mains electrical supply (where appropriate) and switched on. The anaesthetic workstation should be connected directly to the mains electrical supply and only correctly rated equipment connected to its electrical outlets. Multisocket extension leads must not be plugged into the anaesthetic machine outlets or used to connect the anaesthetic machine to the mains supply. Hospitals should have backup generators, and many operating theatres have their own backup system. Backup batteries for anaesthetic machines and other equipment should be charged.
The gas supply master switch (if one is fitted) should be switched on. On some workstations it is necessary to disconnect the oxygen pipeline to check the correct function of the oxygen failure alarm, although on machines with a gas supply master switch, the alarm may be operated by turning the master switch off. Repeated disconnection of gas hoses may lead to premature failure of the Schrader socket and probe, and current guidelines recommend that the regular check of equipment includes a tug test to confirm correct insertion of each pipeline into the appropriate socket (see later). It is also recommended that the oxygen failure alarm is checked once a week by disconnecting the oxygen pipeline with the oxygen flowmeter turned on; the alarm must sound for at least 7 s. Oxygen failure warning devices are also linked to a gas shut-off device. Anaesthetists must be aware of both the tone of the alarm and which gases will continue to flow on the anaesthetic machine.
The suction apparatus should be checked to ensure that it is clean and functioning, all connections are secure and an adequate negative pressure is generated.
The gases that are being supplied by pipeline should be identified, confirming with a tug test that each pipeline is correctly inserted into the appropriate gas supply terminal. Only gentle force is required; excessive force during a tug test may damage the pipeline or gas supply terminal. It is essential to check that the anaesthetic machine is connected to a supply of oxygen and that an adequate reserve supply of oxygen is available from a spare cylinder. It is also necessary to check that adequate supplies of any other gases intended for use are available and connected.
All cylinders should be seated securely and turned off after checking their contents. Carbon dioxide cylinders should not be present on the anaesthetic machine. If a blanking plug is supplied, it should be fitted to any empty cylinder yoke.
All pressure gauges for pipelines connected to the anaesthetic machine should indicate a pressure of 400–500 kPa.
If flowmeters are present, their function should be checked, ensuring that each control valve operates smoothly and the bobbin moves freely throughout its range without sticking. If nitrous oxide is to be used, the hypoxic-guard device should be tested by first turning on the nitrous oxide flow and ensuring that at least 25% oxygen also flows. The oxygen flow should then be turned off to check that the nitrous oxide flow also stops. The oxygen flow should then be turned back on, the nitrous oxide flow should be turned off, and a check made that the oxygen analyser display approaches 100%. All flow control valves should then be turned off; machines fitted with a gas supply master switch will continue to deliver a basal flow of oxygen.
The emergency oxygen bypass control should be operated to ensure that flow occurs from the gas outlet without a significant decrease in the pipeline supply pressure. It is important to ensure that the emergency oxygen bypass control ceases to operate when released; there is a risk of awareness if it continues to operate.
All breathing systems which are to be used must be checked and a two-bag test performed before use ( Box 22.2 ). Breathing systems should be inspected visually for correct configuration and assembly. All connections within the system and to the anaesthetic machine should be checked to ensure that they are secured by push and twist. It should be ensured that there are no leaks or obstructions in the reservoir bags or breathing system. A pressure leak test (between 20 and 60 cmH 2 O) should be performed on the breathing system by occluding the patient end and compressing the reservoir bag.
A two-bag test should be performed after the breathing system, vaporisers and ventilator have been checked individually.
Attach the patient end of the breathing system (including angle piece and filter) to a test lung or bag.
Set the fresh gas flow rate to 5 L min −1 and ventilate manually. Check that the whole breathing system is patent and that the unidirectional valves are moving. Check the function of the adjustable pressure-limiting (APL) valve by squeezing both bags.
Turn on the ventilator to ventilate the test lung. Turn off the fresh gas flow or reduce to a minimum. Open and close each vaporiser in turn. There should be no loss of volume in the system.
Manual leak testing of vaporisers (see later) was previously recommended. It should only be performed on basic Boyle machines because it may be harmful to modern anaesthetic workstations. Manufacturer's recommendations should be reviewed before performing a manual test.
The anaesthetist should check that the vaporisers for the required volatile anaesthetic agents are fitted correctly to the anaesthetic machine, any locking mechanism is fully engaged and the control knobs rotate fully through the full ranges. Vaporisers should be adequately filled but not overfilled, and the filling port must be tightly closed. Tilting vaporisers can result in delivery of dangerously high concentrations of vapour, and therefore they must always be kept upright. All vaporisers must be turned off after they have been checked.
It may be necessary to change a vaporiser during use, although this should be avoided if at all possible. If a change is necessary, repeat the leak test because failure to do so is a common cause of critical incidents. Some anaesthetic workstations automatically test the integrity of vaporisers.
It is only necessary to remove a vaporiser from a machine to refill it if the manufacturer recommends this.
With the vaporiser turned off, a flow rate of oxygen of 5 L min −1 should be set, and the common gas outlet should be temporarily occluded. There should be no leak from any part of the vaporiser, and the flowmeter bobbin (if present) should dip. If more than one vaporiser is present, turn each one on in turn and repeat this test. After the tests, ensure that the vaporisers and flowmeters are turned off.
The contents and connections should be checked to ensure that there is an adequate supply of CO 2 absorbent and that it is of an appropriate colour.
If a coaxial system is in use, an occlusion test should be performed on the inner tube and a check should be made that the adjustable pressure-limiting (APL) valve, where fitted, can be fully opened and closed.
Special care must be exercised if the anaesthetic machine incorporates an auxiliary common gas outlet (ACGO). Incidents of patient harm have resulted from misconnection of a breathing system to an ACGO or incorrect selection of the ACGO.
Whenever a breathing system is changed, either during a case or a list, its integrity and correct configuration must be confirmed. This is particularly important for paediatric cases, when breathing systems may be changed frequently during an operating list.
It is important to check that the ventilator is configured correctly and the ventilator tubing is attached securely. The controls should be set according to the intended use of the ventilator, and the system should be checked to ensure that adequate pressure is generated during the inspiratory phase. Ventilator alarms should be checked to ensure that they are working and correctly configured. The pressure relief valve should be checked to ensure that it functions correctly at the set pressure.
The anaesthetic gas scavenging system should be checked to ensure that it is switched on and functioning and that the tubing is attached to the appropriate exhaust port of the breathing system, ventilator or anaesthetic workstation.
Become a Clinical Tree membership for Full access and enjoy Unlimited articles
If you are a member. Log in here