Conventional/digital subtraction dacryocystography
CT dacryocystography
Magnetic resonance dacryocystography.
Dacryocystography allows visualization of the lacrimal system by direct injection of contrast into the canaliculus of the eyelid. It may be combined with CT imaging performed immediately following lacrimal system contrast injection. CT imaging gives additional diagnostic information, particularly about adjacent structures.
Epiphora – to demonstrate the site and degree of obstruction.
None.
Low osmolar contrast material (LOCM) 300, 0.5–2.0 ml per side.
The use of Lipiodol is to be discouraged.
Digital subtraction radiography unit
Silver punctum or lachrymal dilator and lacrimal probe set
Lacrimal cannula or 18G blunt needle with polythene catheter (the catheter technique has the advantage that the examination can be performed on both sides simultaneously).
0.5% bupivacaine eye drops.
The lacrimal sac is massaged to express its contents prior to injection of the contrast medium. The lower eyelid is everted to locate the lower canaliculus at the medial end of the lid.
The lower punctum normally measures 0.3 mm in diameter and is gently dilated. The lower lid should be drawn laterally to straighten the curve in the canaliculus. To avoid perforation, the cannula is initially positioned perpendicular to the eyelid margin as it enters the punctum. The cannula is then rotated by 90° horizontally along the natural curve of the lower canaliculus.
Care is taken not to advance the cannula too far into the canaliculus as proximal stenosis can be missed.
The upper canaliculus may also be cannulated if there is difficulty with the lower canaliculus.
The contrast medium is injected and a digital subtraction run at one image per second is obtained. This allows for a dynamic study in real time.
Appropriate images are filmed/digitally stored.