Medical Imaging Technologies and Imaging Considerations for 3D Printed Anatomic Models


Medical imaging technologies, which allow us to “see into” and understand living systems, play a significant role in biology and medicine. The X-ray has paved the way for many high-tech medical imaging technologies that are used today including computed tomography (CT), magnetic resonance imaging (MRI), ultrasound, and positron emission tomography (PET). It is possible a number of scientists inadvertently produced X-rays including William Morgan and Nicola Tesla. However, the real discovery and understanding of the X-ray was by Wilhelm Röntgen in late 1895 when he realized X-rays could pass through opaque objects and could be used to generate an image. Röntgen was the first to document this phenomenon using a recording of a photograph of the bones in his wife’s hand, and he coined the term “X-ray.”

X-rays are a penetrating form of electromagnetic radiation carried by photons, with a wavelength ranging from 0.01 to 10 nm that places them into an ionizing radiation category. They are produced in an X-ray tube when a beam of high-energy electrons strikes a high-density target, resulting in production of energy in the form of both heat and these high-energy photons. X-rays have a shorter wavelength and higher frequency than that of visible light.

Not long after X-rays were discovered, the first use of X-rays under clinical conditions was by John Hall-Edwards in Birmingham, England, in 1896 when he radiographed a needle stuck in the hand of an associate. On February 5, 1896, J. T. Bottomley, Lord Blythswood, and John Macintyre gave a presentation to the Philosophical Society of Glasgow on X-rays. Soon after, in March 1896, Mcintyre obtained the permission of the managers of Glasgow Royal Infirmary to set up an X-ray department, which was the first one to be established in the world.

Since the development of the X-ray, there have been many major advancements in medical imaging, and the use of volumetric medical imaging including CT and MRI have played a critical role in the development of medical 3D printing over the years with advancements in both fields being closely intertwined over the decades ( Fig. 2.1 ).

Fig. 2.1
Major early development in medical imaging and 3D printing.

Today, volumetric medical imaging is the backbone of 3D printing in medicine since imaging data are used to create patient-specific, 3D printed anatomic models and guides. Anatomic models may be created from any volumetric dataset, with sufficient contrast and spatial resolution to separate structures, using dedicated image post-processing software (discussed in Chapter 3). When a 3D model is requested prior to an imaging study, the image acquisition and reconstruction techniques should be tailored so that the anatomy in the intended 3D model can be effectively visualized, with the optimal imaging modality, imaging protocol, and image reconstruction depending on the anatomy being imaged.

CT is the most common clinical technique adopted to generate 3D printed anatomic models due to the fact that CT uses X-rays; accordingly, there is only one contrast mechanism, making image post-processing of CT data relatively simple. MRI, which provides exquisite soft tissue contrast without using ionizing radiation, and 3D ultrasound data may also be used. However, implementation is challenging and time-consuming for MRI (due in part to intrinsically low signal-to-noise ratio (SNR) and image artifacts resulting in nonanatomical signal variation); and implementation is even more difficult for ultrasound data due to the non-tomographic nature of ultrasound acquisition. This chapter will review the imaging systems that are typically used to create 3D printed anatomic models including CT, MRI, and ultrasound. Nuclear medicine imaging modalities and utilization of 3D printing will be discussed in Chapter 12.

Computed Tomography

The history of CT dates back to 1917 with the invention of the Radon transform by Johann Radon, an Austrian mathematician, who showed that a function could be reconstructed from an infinite set of its projections. A comprehensive review of the mathematics is available in many texts including Epstein et al. In short, the transformation that takes one-dimensional (1D) projections of a two-dimensional (2D) object over many angles is called the 2D Radon transform. For a fixed angle theta (q) , g(l,q) , is called a projection; and for all l and q , g(l,q) , is the 2D Radon transform of f(x,y) . In order to represent an image, the Radon transform takes multiple, parallel-beam projections of an image from different angles by rotating the source around the center of an image at a specified rotation angle.


R θ ( x ' ) = - f ( x ' cos θ - y ' sin θ , x ' sin θ + y ' cos θ ) d y ' w h e r e x ' y ' = cos θ sin θ - sin θ cos θ x y

The projection of an object results from the tomographic measurement process at any given angle, θ , and is made up of a set of line integrals ( Fig. 2.2 ). The set of projections of a single slice is called a sinogram.

Fig. 2.2, Horizontal and vertical projections for a simple two-dimensional function.

In general, the Radon transform of f(x,y) is the line integral of f parallel to the y’-axis:

The first CT scanner was created in 1967 by Sir Godfrey Hounsfield who, while working at Electrical and Musical Industries (EMI) Central Research Laboratories, used X-rays applied at various angles to create an image of an object. Hounsfield was working on the pattern recognition of letters and was trying to reconstruct a 3D representation of the contents of a box when he realized that it was easier to reconstruct the volume by considering it as a series of slices. Once Hounsfield had verified this principle, he proceeded to construct a testing rig built on the bed of an old lathe that he had from a prior project. This test unit, now known as the “Lathe bed model,” used Americium 95 as its gamma source and had a photon counter as the detector. Initially, it took 9 days to acquire the images and 2.5 h to reconstruct them with an International Computers Limited mainframe computer system.

The first EMI CT scanner was installed in 1971 at the Atkinson Morley Hospital in England and the first human patient brain exam was performed. CT scanners were introduced in the United States (US) a couple of years later with the first clinical scan performed at the Mayo Clinic in 1973. In 1975, EMI marketed the first body scanner which was inaugurally installed at Northwick Park Hospital in London. Shortly after, the first body scanner in the US was installed at the Mallinckrodt Institute at the Washington University School of Medicine in St. Louis, Missouri. By 1976, approximately 17 distinct companies were offering CT scanners and over 475 were sold worldwide. In 1979, Sir Godfrey Hounsfield and Allan McLeod Cormack were both awarded the Nobel Prize in medicine for their pioneering work in the development of CT.

The standard method of reconstructing CT slices is the “backprojection” method which originates from the fact that a 1D projection needs to be filtered by a 1D Radon kernel (backprojected) in order to obtain a 2D signal. The backprojection method starts from a projection value and backprojects a ray of equal pixel values that would sum to the same value ( Fig. 2.3 ). Structures in a complete image can only be restored if a sufficient number of projection angles are collected; however, blurring of the computed image can result if this relatively simple method is employed. Therefore, it is more common to use a filtered backprojection method where a stabilized and discretized version of the inverse Radon transform is typically used.

Fig. 2.3, (A) Projections of point object from three directions and (B) Backprojection onto reconstruction plane.

Since the invention of CT, there have been substantial improvements in speed, slice count, radiation dose, and image quality. The important phases in CT development are single-slice CT, helical CT, multiple detector row CT (MDCT) also known as multislice CT, and spectral CT imaging. Single-slice CT systems have 1D detector arrays, thereby acquiring data in a single section plane and reconstructing only one plane per rotation. Helical CT systems perform acquisitions by continually rotating the source and detectors around the patient as the table is moved along the axis of the scanner, enabling the scan to be performed rapidly ( Fig. 2.4 ). The speed of the table is important for exams such as CT angiography where table motions oscillate at several cm/second.

Fig. 2.4, Illustration of helical CT scanning.

MDCT, introduced in 1992, combines multiple rows of detectors and fast gantry rotation speeds to rapidly gather a cone of X-ray data. MDCT scanners have rapidly evolved from four detector row systems ( Fig. 2.5 ) to systems with 256-slice and 320 detector rows. In comparison to single-row detector CT, MDCT scanners have the advantages of shorter scan durations, longer scan ranges, and thinner slice sections which are especially useful for 3D rendering.

Fig. 2.5, Illustration of MDCT scanning.

Spectral CT imaging uses different energy levels to capture images, which allows elements in the body to be differentiated based on their material density or atomic number. The energy of the reconstructed image can be altered on a sliding scale and can be processed to create many advantages over conventional CT such as the creation of virtual unenhanced images and amplification/removal of iodine or calcium suppressed images. A comprehensive review by Erik Fredenberg lists the many advantages of this acquisition type including improved soft-tissue and contrast agent–based contrast-to-noise ratio (CNR) and better detection of lesions and plaque characterization.

Cone beam computed tomography (or CBCT), a variant type of CT, uses a cone-shaped X-ray beam and 2D detectors instead of a fan-shaped X-ray beam and 1D detectors. CBCT is important in the field of oral and maxillofacial surgery due to the advantages over conventional 2D X-ray methods, such as ease of imaging in a seated or standing position with a rotation of the gantry around the head. With the availability of the equipment in dentist offices, this method is also used in orthodontics and endodontics. Additionally, image-guided radiation therapy often utilizes this method.

CT images are comprised of grayscale pixels, also known as picture elements. The physical size of a pixel depends on the image resolution which is set in the image acquisition. Similar to a pixel, a voxel represents an array of elements of a volume that constitute a 3D image. Both pixels and voxels typically do not have their position/coordinates explicitly encoded with their values. Instead, rendering systems infer the position of pixels and voxels based on their position relative to other pixels/voxels.

In CT, there is only one contrast mechanism; the grayscale intensity of the image is proportional to the tissue density with the grayscale value seen in each pixel reflecting how the amount of the X-ray beam the tissue attenuated as it passed through the patient. The Hounsfield unit (HU) scale, named after Sir Godfrey N. Hounsfield, used in these images is a linear transformation of the original linear attenuation coefficient mechanism, and it provides information regarding what type of tissue types may be present (i.e., air, bone, soft tissue, blood, etc.). On the HU scale, air is represented by a value of −1000 and appears black on the grayscale image and bone is +1000 and appears white on the grayscale image ( Fig. 2.6 ).

Fig. 2.6, Example coronal CT image of the chest, abdomen, and pelvis with Hounsfield Units (HUs) for certain tissues shown including the lungs (air) with a mean HU = –735.45, aorta (blood pool) with a mean HU = 452.65, kidney with a mean HU = 320.35, and pelvis (bone) with a mean HU = 1048.91.

Magnetic Resonance Imaging

MRI is a dynamic and flexible noninvasive imaging technology that operates at the radiofrequency (RF) range and produces detailed anatomic images without ionizing radiation exposure. MRI was originally called NMRI (nuclear magnetic resonance imaging) after the nuclear magnetic resonance (NMR) phenomenon, first described by Isidor Isaac Rabi in 1938. By passing a particle beam of lithium chloride molecules through a vacuum chamber and manipulating the beam with different magnetic fields, Rabi demonstrated how the magnetic moments of nuclei could be induced to flip their principal magnetic orientation. Rabi won the Nobel Prize in Physics for this discovery in 1944. With the NMR foundation in hand, Felix Bloch at Stanford and Edward Mills Purcell at Harvard measured the precession of spins around a magnetic field in condensed matter, using a concept which is known today as continuous-wave NMR. , Bloch and Purcell were awarded the 1952 Nobel Prize for Physics for their development of new approaches and methods for nuclear magnetic precession measurements.

The physical basis of NMR centers around the concept of a nuclear spin, which gives a nucleus a magnetic moment related to its angular momentum by


μ = γ s ,

where μ is the magnetic moment, γ is the gyromagnetic ratio (a nucleus-dependent property), and s is the spin angular momentum. The NMR effect can only be observed for nuclei with odd numbered protons or neutrons such as 1 H (protons), 11 Na, 13 C, 19 F, or 31 P. Since hydrogen atoms ( 1 H), with a nucleus consisting of a single proton, are most prevalent in human tissue, most MRI techniques focus only on the resonance effects of these protons. For 1 H, the gyromagnetic ratio is 42.58 MHz/T.

To measure nuclear resonance, a sample is placed in a static magnetic field, B 0 . In this field, the magnetic moment vectors of individual nuclei align with the direction of the field, and the protons precess at a well-defined frequency based on the field strength, called the Larmor frequency ( ω 0 ) which is proportional to B 0 and is defined as


ω 0 = γ B 0

At rest, nuclei precess about B 0 , but since they are all randomly oriented with respect to each other, there is no detectable signal. When external energy in the form of a RF magnetic field is applied at the Larmor frequency, resonance, a natural phenomenon characterized by an oscillating response occurs. The 1 H nuclei are excited out of their equilibrium state into a state of alignment tipped away from the direction of B 0 ; however, they still precess about B 0 , and their precessional motion results in their net nuclear magnetic fields inducing RF currents in nearby conductors, or RF coils, emitting a detectable signal. In the human body, signals vary depending on the tissue type with greater proton density yielding to a higher signal and vice versa. Depending on the surrounding molecules (i.e., tissue type), 1 H nuclei within the magnetic field return to a state of equilibrium at different rates. If the spins aligned with B 0 are added together, they result in a net magnetization vector ( Fig. 2.7 ).

Fig. 2.7, Illustration of the alignment of spins in magnetic field and precession. Schematic representation of a group of microscopic magnetic moments (Left), Alignment of magnetic moments in the presence of an external magnetic field (Middle), Net magnetization vector (Right).

In NMR, free induction decay (FID), first performed by Erwin Hahn, is defined as the observable NMR signal generated by non-equilibrium nuclear spin magnetization precessing around B 0 . In FID, following RF excitation, the magnetization relaxes and the protons undergo dephasing which attenuates the transverse component of the magnetization and causes the signal to decay and energy loss which causes the longitudinal component of the magnetization to recover its equilibrium value. The magnitude of a FID signal is dependent upon a number of parameters such as the sample size and composition, the strength of the magnetic field, and the flip angle (FA). The FA, also known as the tip angle, is the amount of rotation the net magnetization (M) experiences during the application of an RF pulse ( Fig. 2.8 ).

Fig. 2.8, Illustration showing the effect of an RF pulse (Left) and subsequent recovery after RF pulse is turned off (Right).

In MRI, after excitation, each tissue returns to its equilibrium state. Longitudinal or spin-lattice relaxation describes the loss of energy from the spins to their environment, causing recovery of the longitudinal component of the equilibrium value M 0 , with time constant T 1. Transverse or spin–spin relaxation is the irreversible dephasing among spins due to microscopic interactions between molecules and nuclei in the transverse direction; it occurs with time constant T 2 . The choice of echo time (TE: the time between the initial excitation pulse and the time at which signal is acquired) and repetition time (TR: the time between successive acquisitions, equivalent to the time between successive excitations) determines the signal contrast. If a short TR is chosen, tissues with long T 1 do not fully recover to equilibrium between successive excitations, thus they produce a weak signal. Alternatively, tissues with short T 1 recover more quickly and produce a relatively strong signal. Table 2.1 shows relaxation times at 3T for some sample tissues. Different pulse sequences may be exploited to produce signal contrast in MRI and these MRI techniques may be adapted to optimize conspicuity of certain structures or pathologies.

Table 2.1
T 1 and T 2 Relaxation Times for Select Tissues at 3T.
Derived from de Bazelaire CM, Duhamel GD, Rofsky NM, Alsop DC. MR imaging relaxation times of abdominal and pelvic tissues measured in vivo at 3.0 T: preliminary results. Radiology . 2004;230(3):652–659.
T 1 (ms) T 2 (ms)
Kidney cortex 1142 ± 154 76 ± 7
Kidney medulla 1545 ± 142 81 ± 8
Spleen 1328 ± 31 61 ± 9
Paravertebral muscle 898 ± 33 29 ± 4
Subcutaneous fat 382 ± 13 68 ± 4
Prostate 1597 ± 42 74 ± 9
Pancreas 725 ± 71 43 ± 7
Cervix 1616 ± 61 83 ± 7

The concept of pulse sequences was first introduced by Erwin Hahn who demonstrated that the NMR effect could also be observed with a modified experimental method based on finite RF pulses. Hahn also discovered the spin echo which results from two successive RF pulses, a 180 degree pulse following a 90 degree pulse, representing regeneration of spin phase information lost during the decay of the FID. In a conventional spin echo pulse sequence, one line of imaging data is collected within each TR period and the pulse sequence is repeated for multiple TR periods until all phase-encoding steps are collected. By altering the echo times, MRI pulse sequences may be tailored to test how fast these protons return to equilibrium, which influences the overall appearance of the signal, thus highlighting certain anatomy. Though this is an excellent method to produce soft tissue contrast, it does not present a singular contrast mechanism, like CT, which can affect post-processing protocols.

Magnetic resonance spectroscopy was proposed as a means of differentiating tumor and healthy tissues for the first time in 1964 by John Mallard. Subsequently, Raymond Damadian discovered that there was a marked difference in the T 1 and T 2 times between normal and abnormal tissues, suggesting that NMR relaxation times could be used to distinguish cancer from healthy tissue. This work was published in 1971. In 1973, Paul Lauterbur was the first to demonstrate that it was possible to use NMR to create an image using linear gradient fields to distinguish between NMR signals originating from different locations and combined this with a form of reconstruction. That same year, Sir Peter Mansfield published his method for the determination of spatial structures in solids by NMR which formed the basis of slice selection used routinely until today. In 1974, James Hutchison, from the University of Aberdeen, likely showed the first pathology with MRI using projection reconstruction—the image was colored using pencils but clearly showed the rat’s cause of death—cervical dislocation (see Fig. 2.9 ).

Fig. 2.9, This is likely the first MRI image demonstrating pathology. A dead mouse was imaged on a 0.06T MRI scanner with projection-reconstruction imaging based upon 25 projections using an inversion recovery sequence which generated a T 1 image. The array of numbers was colored by hand. Blood pooling around the neck injury is colored in black.

The echo-planar imaging (EPI) technique which allowed multiple lines of imaging data to be acquired after a single RF excitation was developed by Mansfield in 1975. Similar to conventional spin echo, spin echo EPI starts out with a 90 degree pulse followed by a 180 degree pulse. Following the 180 degree pulse, the frequency-encoding gradient oscillates rapidly from a positive to a negative amplitude, forming a train of gradient echoes. By using a resonant gradient system, images could be acquired in 100 ms or less during the rapid oscillation of a magnetic field gradient. Using this technique, Mansfield produced the first MR image of live human anatomy.

In the late 1970s and early 1980s, the University of Aberdeen team worked to make MRI useable and robust by combining the concepts of slice selection, frequency encoding, and phase encoding in the Spin Warp Imaging sequence which became the foundation of MRI sequences and enabled the widespread adoption and industrialization of MRI for the decades to come ( Fig. 2.10 ). The phrase “Spin Warp” actually was conceived by Jim Hutchison’s wife Margaret Foster (a highly respected scientist in her own right) who was an avid fan of Star Trek. Together, Sir Peter Mansfield and Paul Lauterbur were awarded the 2003 Nobel Prize in Physiology and Medicine for their “discoveries concerning MRI.”

Fig. 2.10, Spin warp imaging sequence diagram.

Following Aberdeen’s success, the first commercial MR imaging systems were introduced in the early 1980s. These systems had limited options for pulse sequences, typically having only basic spin-echo and gradient-echo techniques allowing for T 1 - and T 2 -weighted imaging. Over the past 40 years, there have been significant advancements in hardware, pulse sequences, and image reconstruction algorithms, ultimately allowing for better and faster MR imaging. The use of superconducting magnets has enabled imaging at high field strengths; today, 1.5 and 3T magnets are routinely used for clinical imaging. Improved speed and resolution of MRI examination have also been achieved with use of phased array coil detection, parallel imaging, and sensitivity encoding. More recently, MRI has moved from being a tool for seeing pure morphology to one of understanding basic physiology.

Current MRI Systems

In modern-day MRI, a slice can be prescribed in any plane and with any slice thickness. Thin slices are generally prescribed in 2D imaging, but for 3D imaging thick slabs are typically utilized. The center frequency of the RF pulse controls the location of the imaging plane along the direction of the slice select gradient. For a slice at isocenter, the scanner applies an RF pulse at the center frequency. If slices far away from isocenter need to be collected, an RF pulse with a frequency offset must be applied. An RF pulse of finite duration will have a band of frequencies, called the bandwidth (BW). The slice thickness Δz can be described in terms of the RF pulse BW and an applied gradient in z-direction Gz,


Δ z = BW / γ G z .

Image reconstruction techniques in MRI are dependent on how the spatial information (frequency and phase) is encoded. The basic principal of frequency encoding is that an applied gradient causes nuclear spins at different locations in the direction of that gradient to precess at different frequencies as the signal is acquired. The phase-encoding gradient is a magnetic field gradient that allows the encoding of a spatial signal location. The frequency and phase of the signal are encoded by the modulation of gradient fields in space and time; and this relationship can be determined by the Fourier transform.

For the purposes of MRI, the 1D Fourier transform plays a very important role in allowing a signal to be decomposed into its frequency components written as


S ( k x ) = I ( x ) e i 2 π k x x d x

where S is the detected signal at one point in time, I is the amplitude of the signal source (which is a function of the relaxation times T 1 , and T 2 and proton density at each location in space as well as on TR and TE of the pulse sequence and numerous other parameters), x is location in space, and k x is the immediate location in k-space, which is a function of the duration and strength of the field gradient orientated in x .

The use of a Fourier transform to process these signals would appear as a single projection through the excited slice of the object in question. Simple backprojections could then be used to produce an image as shown by Lauterbur, but other CT-type Radon Transform reconstruction techniques could also be utilized. It is much more common though to acquire all the data in the multidimensional spatial frequency-domain “k-space” and generate an image by performing a multidimensional Fourier transform. K-space is therefore the mathematical space representing the Fourier transform of a spatial function.

K-space data that are sampled rectilinearly can more easily be reconstructed with Fourier transformation in all orthogonal spatial dimensions without any prior weighting or interpolation. Rectilinear data acquisition uses phase encoding in combination with frequency encoding to resolve the locations of the spins in both in-plane directions ( Fig. 2.11A ). Phase-encoding gradients, which are applied for finite durations, produce a unique phase in the signal which is a linear function of position in the direction that is conventionally orthogonal to that used for frequency encoding. Sampling of k-space may also be performed radially, with frequency-encoding gradients applied sequentially in many different directions as described above ( Fig. 2.11B ). For radial acquisitions, a 2D Fourier transform can only produce a 2D image of the slice after special interpolation or weighting of the data points is done due to their uneven distribution in k-space.

Fig. 2.11, Illustration of the (A) Cartesian sampling scheme (B) and radial scheme.