Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
To adequately prepare the target lesion to facilitate stent delivery and expansion.
( Fig. 9.1 )
Lesion preparation (in most cases with balloon angioplasty) should be performed in nearly all lesions because it:
Facilitates stent delivery and decreases the risk of stent loss.
Helps determine optimal stent diameter and length (especially when no intracoronary imaging is used and when there is poor flow of contrast distal to the target lesion).
Facilitates stent expansion and helps determine the need for additional lesion modification (e.g., with atherectomy in heavily calcified lesions) to prevent “stent-regret.”
Coronary lesion preparation should not be performed until after successful wiring of the lesion (as described in Chapter 8 : Wiring). Optimal distal wire position should be confirmed before balloon angioplasty (and/or stenting).
If large thrombus is present, coronary thrombectomy is often performed, as described in Chapter 20 : Acute coronary syndromes—thrombus.
If severe calcification is present, atherectomy or intracoronary lithotripsy is often performed, as described in Chapter 19 : Calcification.
Primary stenting (i.e., stenting without balloon angioplasty or other lesion preparation) may be preferred in few lesion types, such as degenerated saphenous vein grafts with friable lesions, or lesion with thrombus (when no thrombectomy is planned or when thrombectomy fails).
To ensure that the guidewire is optimally positioned (through the target lesion and advanced several centimeters distally, but not into small distal branches that could lead to perforation).
Contrast injection is performed to confirm that the guidewire is in optimal position. If position is suboptimal, the guidewire is repositioned.
If the guidewire is suboptimally positioned, complications can happen at the time of balloon inflation, as follows:
Causes:
Wire entered small distal branch.
Prevention:
Ensure optimal distal guidewire position (wire should not be in a small branch to minimize the risk of distal coronary perforation; this is, especially important for branches located close to the occlusion, as the balloon may enter that branch and perforate it while dilating the lesion).
Treatment:
Treatment of perforation is discussed in Chapter 26 : Perforation.
To ensure there is no air or debris inside the guide catheter.
Back bleed the Y-connector.
Aspirate the guide catheter, then fill with contrast.
Causes:
Air sucked into the guide catheter during balloon withdrawal. This is more likely when withdrawing large balloons.
Thrombus or plaque entry into the guide catheter while withdrawing a balloon (may be more likely with bulky large diameter balloons, or plaque modification balloons, such as the Angiosculpt, Chocolate, and cutting balloon, Section 30.9.3 ).
Prevention:
Guide aspiration followed by flushing, as described above.
Treatment:
Air embolization is treated with administration of 100% oxygen, possibly aspiration, and administration of intracoronary epinephrine in case of cardiac arrest. Air embolization usually resolves with supportive measures without needing additional intervention. Air embolization is described in detail in Section 25.2.3.3 .
Embolization of plaque or thrombus is usually treated with thrombectomy, as described in Chapter 20 : Acute coronary syndromes—thrombus.
To choose optimal balloon size and type.
For achieving lesion expansion : for most lesions the balloon diameter is chosen to match the distal reference vessel diameter (1:1 ratio). The goal is to determine whether the balloon fully expands and therefore no additional lesion modification is needed before stenting. This is particularly important for calcified and in-stent restenotic lesions.
For crossing : for very tight lesions that are hard to cross, small balloons (≤1.5 mm, Section 30.9.2 ) are initially used to modify the lesion entry point and allow subsequent delivery of larger balloons to further modify the lesion before stent placement.
Balloon length should be shorter than the estimated lesion length (to avoid injury of coronary segments proximal or distal to the target lesion that will require implantation of longer length stents).
Noncompliant balloons are preferred, because they can also be used for postdilation of the target lesion after stenting. However, they are less deliverable than compliant balloons.
There are two balloon delivery systems: monorail and over-the-wire. Monorail balloons should be used in the vast majority of cases, as they are simpler to use, do not require use of long guidewires, and are easier to deliver because they have a stiffer shaft.
Some balloons have nitinol wires (such as the Angiosculpt and the Chocolate) or cutting blades (cutting balloon) aiming to modify the lesion and facilitate expansion ( Section 30.9.3 ). Plaque modification balloons are harder to deliver (due to larger profile and lesser flexibility). The cutting balloon also requires slow inflation and deflation. Plaque modification balloons can, in some cases, facilitate expansion of “balloon undilatable lesions” ( Section 23.2 ). The SIS OPN balloon ( Section 30.9.7 ) can be very useful for dilating resistant lesions, but is not currently available in the United States.
Remove the air from the balloon, fill the balloon with a contrast solution, and connect to an indeflator.
The balloon is removed from its packaging.
A luer-lock syringe is filled usually with a 50% contrast/50% saline solution. Lower concentration of contrast will speed the inflation/deflation of the balloon but will decrease the balloon visibility.
The syringe is connected with the balloon proximal hub.
Negative suction is performed with the syringe plunger positioned up.
The balloon is connected with the indeflator and negative suction is applied.
Operator finger injury . The balloon/stent should be slowly removed from the container hoop to reduce the risk of the balloon/stent stylet injuring the operator’s hands.
Causes:
Inadvertent filling of the balloon with saline instead of contrast solution.
Poor balloon preparation with significant amount of air remaining in the balloon.
Prevention:
Ensure that diluted contrast (and not pure saline) is used to prepare and inflate the balloon. The syringe containing the contrast/saline solution should be appropriately labeled.
Ensure that air is removed during balloon preparation.
Treatment:
If balloon cannot be visualized once inflated, it should be prepared again using a contrast solution and optimal preparation technique.
To load the balloon on guidewire to allow subsequent advancement to the target coronary lesion.
Become a Clinical Tree membership for Full access and enjoy Unlimited articles
If you are a member. Log in here